G1312-90008_binarypumpvl_user_en.fm
Agilent 1260 Infinity
Binary Pump VL
Agilent Technologies, Inc. 2008,
receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
The material contained in this docu-
1987). U.S. Government users will receive
No part of this manual may be reproduced
ment is provided "as is," and is sub-
no greater than Limited Rights as defined in
in any form or by any means (including elec-
ject to being changed, without notice,
FAR 52.227-14 (June 1987) or DFAR
tronic storage and retrieval or translation
in future editions. Further, to the max-
252.227-7015 (b)(2) (November 1995), as
into a foreign language) without prior agree-
imum extent permitted by applicable
applicable in any technical data.
ment and written consent from Agilent Technologies, Inc. as governed by United
law, Agilent disclaims all warranties,
States and international copyright laws.
either express or implied, with regard
to this manual and any information
Manual Part Number
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a par-
A CAUTION notice denotes a
ticular purpose. Agilent shall not be
hazard. It calls attention to an
liable for errors or for incidental or
operating procedure, practice, or
consequential damages in connection
with the furnishing, use, or perfor-
the like that, if not correctly per-
Printed in Germany
mance of this document or of any
formed or adhered to, could
Agilent Technologies
information contained herein. Should
result in damage to the product
Hewlett-Packard-Strasse 8
Agilent and the user have a separate
or loss of important data. Do not
written agreement with warranty
proceed beyond a CAUTION
This product may be used as a com-
terms covering the material in this
ponent of an in vitro diagnostic sys-
document that conflict with these
notice until the indicated condi-
tem if the system is registered with
terms, the warranty terms in the sep-
tions are fully understood and
the appropriate authorities and com-
arate agreement shall control.
plies with the relevant regulations.
Otherwise, it is intended only for gen-
Technology Licenses
eral laboratory use.
The hardware and/or software described in
this document are furnished under a license and may be used or copied only in accor-
A WARNING notice denotes a
dance with the terms of such license.
hazard. It calls attention to an
Restricted Rights Legend
operating procedure, practice,
or the like that, if not correctly
If software is for use in the performance of a U.S. Government prime contract or subcon-
performed or adhered to, could
tract, Software is delivered and licensed as
result in personal injury or
"Commercial computer software" as
death. Do not proceed beyond a
defined in DFAR 252.227-7014 (June 1995), or as a "commercial item" as defined in FAR
WARNING notice until the indi-
2.101(a) or as "Restricted computer soft-
cated conditions are fully under-
ware" as defined in FAR 52.227-19 (June
stood and met.
1987) or any equivalent agency regulation or contract clause. Use, duplication or dis-closure of Software is subject to Agilent Technologies' standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will
1260 Binary Pump VL User Manual
In This Guide.
In This Guide.
This manual covers the Agilent 1260 Infinity Quaternary Pump VL (G1311C).
This chapter gives an introduction to the module, instrument overview and internal connectors.
This chapter provides information on environmental requirements, physical and performance specifications.
This chapter gives information about the preferred stack setup for your system and the installation of your module.
This chapter provides information for optimized usage of the binary pump.
This chapter gives hints on how to optimize the performance or use additional devices.
Overview about the troubleshooting and diagnostic features.
This chapter describes the meaning of error messages, and provides information on probable causes and suggested actions how to recover from error conditions.
1260 Binary Pump VL User Manual
In This Guide.
This chapter describes the tests for the module.
This chapter describes the maintenance of the module.
This chapter provides information on parts for maintenance.
This chapter provides information on cables used with the Agilent 1200 Infinity Series modules.
This chapter describes the pump in more detail on hardware and electronics.
This chapter provides addition information on safety, legal and web.
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
This chapter gives an introduction to the module, instrument overview and internal connectors.
Introduction
Introduction to the Binary Pump
Introduction to the Binary Pump
The binary pump comprises two identical pumps integrated into one housing. It provides gradient generation by high-pressure mixing. Degassing is not included but a vacuum degasser is available as a separate product for applications that require best flow stability especially at low flow rates or maximum detector sensitivity. This is most likely required to run small internal diameter columns (2 mm and 1 mm i.d.) which require low flow rates. A solvent selection valve (optional) will allow to select a binary mixture (isocratic and gradient) from four independent solvent bottles. An active seal wash (optional) is available when the pump is used with concentrated buffer solutions.
Active inlet valve
Overview of the binary pump
1260 Binary Pump VL User Manual
Overview of the Binary Pump
Overview of the Binary Pump
The binary pump is based on a two-channel, dual-piston in-series design which comprises all essential functions that a solvent delivery system has to fulfill. Metering of solvent and delivery to the high-pressure side are performed by two pump assemblies which can generate pressure up to 400 bar.
Each channel comprises a pump assembly including pump drive, pump head, active inlet valve which has a replaceable cartridge, and outlet valve. Both channels are connected in a low-volume mixing chamber which is connected by a capillary coil to a damping unit and a mixer. A purge valve including a PTFE frit is fitted at the pump outlet for convenient priming of the pumping system.
A seal wash (optional) is available when the pump is used with buffer solutions.
1260 Binary Pump VL User Manual
Introduction
Overview of the Binary Pump
The Hydraulic Path
1260 Binary Pump VL User Manual
Overview of the Binary Pump
How Does the Binary Pump Work?
The liquid runs from the solvent reservoir through an active inlet valve. Each side of the binary pump comprises two substantially identical pump units. Both pump units comprise a ball-screw drive and a pump head with two sapphire pistons for reciprocating movement.
Idb m c X]VbWZg
Bdidgl i]ZcXdYZg
Principle of Pump Unit
1260 Binary Pump VL User Manual
Introduction
Overview of the Binary Pump
A servo-controlled variable reluctance motor drives the two ball-screw drives in opposite directions. The gears for the ball-screw drives have different circumferences (ratio 2:1) allowing the first piston to move at double the speed of the second piston. The solvent enters the pump heads close to the bottom limit and leaves it at its top. The outer diameter of the piston is smaller than the inner diameter of the pump-head chamber allowing the solvent to fill the gap in between. The first piston has a stroke volume in the range of 20 µL to 100 µL depending on the flow rate. The microprocessor controls all flow rates in a range of 1 µL/min to 5 mL/min. The inlet of the first pumping unit is connected to the active inlet valve which is processor-controlled opened or closed allowing solvent to be drawn into the first pump unit.
The outlet of the pump unit is connected directly to the second pump unit. The outlet of the second pump unit is connected via a small mixing chamber, a coil and the damping unit to the purge valve assembly. The outlet of the purge valve assembly is then connected to the following chromatographic system.
When turned on, the pump runs through an initialization procedure to determine the upper dead-center of the first piston of both pump channels. The first piston moves slowly upwards to the mechanical stop of the pump head and from there it moves back a predetermined path length. The controller stores this piston position in memory. After this initialization the pump starts operation with the set parameters for the two pump channels.
The active inlet valve is opened and the down moving piston draws solvent into the first pump head. At the same time the second piston is moving upwards delivering into the system. After a controller-defined stroke length (depending on the flow rate) the drive motors are stopped and the active inlet valve is closed. The motor direction is reversed and moves the first piston up until it reaches the stored upper limit and at the same time moving the second piston downwards.
Then the sequence starts again moving the pistons up and down between the two limits. During the upward movement of the first piston the solvent in the pump head is pressed through the outlet ball valve into the second pumping unit. The second piston draws in half of the volume displaced by the first piston and the remaining half volume is directly delivered into the system. During the drawing stroke of the first piston, the second piston delivers the drawn volume into the system.
1260 Binary Pump VL User Manual
Overview of the Binary Pump
From mixing point to pump outlet, dependent on back pressure (180 – 480 µL without mixer, 600 – 900 µL with mixer)
Materials in contact with mobile phase
SST, gold, sapphire, ceramic
Active inlet valve
SST, sapphire, ruby, ceramic, PTFE
Outlet ball valve
SST, gold, sapphire, ruby, tantalum
SST, gold, PTFE, ceramic
For pump specifications, see
How Does Compressibility Compensation Work?
The compressibility of the solvents in use will affect retention-time stability when the back pressure in the system changes (for example, ageing of column). In order to minimize this effect, the pump provides a compressibility compensation feature which optimizes the flow stability according to the solvent type. The compressibility compensation is set to a default value and can be changed through the user interface.
Without a compressibility compensation the following will happen during a stroke of the first piston. The pressure in the piston chamber increases and the volume in the chamber will be compressed depending on backpressure and solvent type. The volume displaced into the system will be reduced by the compressed volume.
With a compressibility value set the processor calculates a compensation volume, that is depending on the backpressure in the system and the selected compressibility. This compensation volume will be added to the normal stroke volume and compensates the previous described loss of volume during the delivery stroke of the first piston.
1260 Binary Pump VL User Manual
Introduction
Overview of the Binary Pump
How Does Variable Stroke Volume Work?
Due to the compression of the pump-chamber volume each piston stroke of the pump will generate a small pressure pulsation, influencing the flow stability of the pump. The amplitude of the pressure pulsation depends mainly on the stroke volume and the compressibility compensation for the solvent in use. Small stroke volumes generate pressure pulsations of smaller amplitude than higher stroke volumes at the same flow rate. In addition, the frequency of the pressure pulsations is higher. This decreases the influence of flow pulsations on quantitative results.
In gradient mode smaller stroke volumes result in a lower flow ripple improve composition ripple.
The module uses a processor-controlled spindle system for driving its pistons. The normal stroke volume is optimized for the selected flow rate. Small flow rates use a small stroke volume while higher flow rates use a higher stroke volume.
By default, the stroke volume for the pump is set to AUTO mode. This means that the stroke is optimized for the flow rate in use. A change to larger stroke volumes is possible but not recommended.
1260 Binary Pump VL User Manual
Early Maintenance Feedback
Early Maintenance Feedback
Maintenance requires the exchange of components which are subject to wear
or stress. Ideally, the frequency at which components are exchanged should be
based on the intensity of usage of the module and the analytical conditions,
and not on a predefined time interval. The early maintenance feedback (EMF)
feature monitors the usage of specific components in the instrument, and
provides feedback when the user-selectable limits have been exceeded. The
visual feedback in the user interface provides an indication that maintenance
procedures should be scheduled.
EMF Counters
EMF counters increment with use and can be assigned a maximum limit which
provides visual feedback in the user interface when the limit is exceeded.
Some counters can be reset to zero after the required maintenance procedure.
Using the EMF Counters
The user-settable EMF limits for the EMF Counters enable the early maintenance
feedback to be adapted to specific user requirements. The useful maintenance
cycle is dependent on the requirements for use. Therefore, the definition of the
maximum limits need to be determined based on the specific operating
conditions of the instrument.
Setting the EMF Limits
The setting of the EMF limits must be optimized over one or two maintenance
cycles. Initially the default EMF limits should be set. When instrument
performance indicates maintenance is necessary, take note of the values
displayed by the EMF counters. Enter these values (or values slightly less than
the displayed values) as EMF limits, and then reset the EMF counters to zero.
The next time the EMF counters exceed the new EMF limits, the EMF flag will be
displayed, providing a reminder that maintenance needs to be scheduled.
1260 Binary Pump VL User Manual
The industrial design of the module incorporates several innovative features. It uses Agilent's E-PAC concept for the packaging of electronics and mechanical assemblies. This concept is based upon the use of expanded polypropylene (EPP) layers of foam plastic spacers in which the mechanical and electronic boards components of the module are placed. This pack is then housed in a metal inner cabinet which is enclosed by a plastic external cabinet. The advantages of this packaging technology are:
• virtual elimination of fixing screws, bolts or ties, reducing the number of
components and increasing the speed of assembly/disassembly,
• the plastic layers have air channels molded into them so that cooling air can
be guided exactly to the required locations,
• the plastic layers help cushion the electronic and mechanical parts from
physical shock, and
• the metal inner cabinet shields the internal electronics from
electromagnetic interference and also helps to reduce or eliminate radio frequency emissions from the instrument itself.
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
2
Site Requirements and Specifications
This chapter provides information on environmental requirements, physical and performance specifications.
Site Requirements and Specifications
Site Requirements
A suitable environment is important to ensure optimal performance of the instrument.
The module power supply has wide ranging capability. It accepts any line voltage in the range described iny there is no voltage selector in the rear of the module. There are also no externally accessible fuses, because automatic electronic fuses are implemented in the power supply.
Hazard of electrical shock or damage of your instrumentation
can result, if the devices are connected to a line voltage higher than specified.
➔ Connect your instrument to the specified line voltage only.
The module is partially energized when switched off, as long as the power cord is
plugged in.
Repair work at the module can lead to personal injuries, e.g. electrical shock, when
the cover is opened and the module is connected to power.
➔ Always unplug the power cable before opening the cover.
➔ Do not connect the power cable to the instrument while the covers are removed.
1260 Binary Pump VL User Manual
Site Requirements and Specifications
Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument from the power line at any time.
➔ Make sure the power connector of the instrument can be easily reached and
➔ Provide sufficient space behind the power socket of the instrument to unplug the
Power Cords
Different power cords are offered as options with the module. The female end of all power cords is identical. It plugs into the power-input socket at the rear. The male end of each power cord is different and designed to match the wall socket of a particular country or region.
Absence of ground connection or use of unspecified power cord
The absence of ground connection or the use of unspecified power cord can lead to
electric shock or short circuit.
➔ Never operate your instrumentation from a power outlet that has no ground
➔ Never use a power cord other than the Agilent Technologies power cord designed
for your region.
Use of unsupplied cables
Using cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
➔ Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
1260 Binary Pump VL User Manual
Site Requirements and Specifications
Site Requirements
Unintended use of supplied power cords
Using power cords for unintended purposes can lead to personal injury or damage of
electronic equipment.
➔ Never use the power cords that Agilent Technologies supplies with this instrument
for any other equipment.
Bench Space
The module dimensions and weight (see low you to place the module on almost any desk or laboratory bench. It needs an additional 2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches) in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Condensation within the module
Condensation will damage the system electronics.
➔ Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
➔ If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.
1260 Binary Pump VL User Manual
Site Requirements and Specifications
Physical Specifications
15.5 kg (34 lbs)
(height × width × depth)
(7 x 13.5 x 17 inches)
100 – 240 VAC, ± 10 %
Wide-ranging capability
Power consumption
220 VA, 74 W / 253 BTU
Ambient operating
4–55 °C (41–131 °F)
Ambient non-operating
-40 – 70 °C (-4 – 158 °F)
< 95 %, at 25 – 40 °C (77 – 104 °F)
Operating altitude
Up to 2000 m (6562 ft)
Non-operating altitude
Up to 4600 m (15091 ft)
For storing the module
Safety standards:
Installation category II, Pollution degree 2
For indoor use only.
1260 Binary Pump VL User Manual
Site Requirements and Specifications
Performance Specifications
Performance Specification Agilent 1260 Infinity Binary Pump VL (G1312C)
Two dual piston in series pumps with servo-controlled variable stroke drive, floating pistons
Setable flow range
Set points 0.001 – 5 mL/min, in 0.001 mL/min increments
0.05 – 5.0 mL/min
≤0.07 % RSD, or ≤0.02 min SD whatever is greater, based on retention time at constant room temperature
±1 % or 10 µL/min whatever is greater, pumping degassed H2O at 10 MPa (100 bar)
Pressure operating range
Operating range up to 40 MPa (400 bar, 5880 psi) up to 5 mL/min
Pressure pulsation
< 2 % amplitude (typically < 1.3 %), or < 0.3 MPa (3 bar) whatever is greater, at 1 mL/min isopropanol, at all pressures > 1 MPa (10 bar, 147 psi)
User-selectable, based on mobile phase
Recommended pH range
1.0 – 12.5 , solvents with pH < 2.3 should not contain acids which attack stainless steel
Gradient formation
High-pressure binary mixing
600 – 900 µL (includes 400 µL mixer), dependent on back pressure
1260 Binary Pump VL User Manual
Site Requirements and Specifications
Performance Specification Agilent 1260 Infinity Binary Pump VL (G1312C)
Composition range
settable range: 0 – 100 %recommended range: 1 – 99 % or 5 µL/min per channel, whatever is greater
Composition precision
< 0.15 % RSD or < 0.04 min SD, whatever is greater, at 0.2 and 1 mL/min; based on retention time at constant room temperature
Composition accuracy
± 0.5 % absolute
(water/caffeine tracer)
Agilent control software (e.g. ChemStation, EZChrom, OL, MassHunter)
Agilent Instant Pilot
For pressure monitoring, 2 mV/bar, one output
Controller-area network (CAN), RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional
Safety and maintenance
Extensive diagnostics, error detection and display through Agilent LabAdvisor, leak detection, safe leak handling, leak output signal for shutdown of the pumping system. Low voltage in major maintenance areas.
Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of seal wear and volume of pumped mobile phase with pre-defined and user settable limits and feedback messages. Electronic records of maintenance and errors
All materials are recyclable
For use with flow rates below 500 µL/min a vacuum degasser is required.
1260 Binary Pump VL User Manual
Site Requirements and Specifications
Performance Specifications
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
3
Installing the Pump
This chapter gives information about the preferred stack setup for your system and the installation of your module.
Installing the Pump
Unpacking the Binary Pump
Unpacking the Binary Pump
If the delivery packaging shows signs of external damage, please call your Agilent Technologies sales and service office immediately. Inform your service representative that the instrument may have been damaged during shipment.
"Defective on arrival" problems
If there are signs of damage, please do not attempt to install the module. Inspection by Agilent is required to evaluate if the instrument is in good condition or damaged.
➔ Notify your Agilent sales and service office about the damage.
➔ An Agilent service representative will inspect the instrument at your site and
initiate appropriate actions.
1260 Binary Pump VL User Manual
Installing the Pump
Unpacking the Binary Pump
Ensure all parts and materials have been delivered with the pump. For checking the completeness of your specific shipment, please use the list included in your shipment. To aid in parts identification, please refer tolease report missing or damaged parts to your local Agilent Technologies sales and service office.
Agilent 1260 Infinity Binary Pump VL
HPLC System Tool Kit (OPTIONAL)
HPLC Starter Kit incl. 0.17 mm i.d. cap (OPTIONAL)
HPLC Starter Kit incl. 0.12 mm i.d. cap (OPTIONAL)
Interface board (LAN) (OPTIONAL)
Agilent 1200 Infinity Series User Documentation DVD (OPTIONAL)not orderable (OPTIONAL)
Bottle-head assembly
Column Eclipse Plus C18, 4.6 x 100 mm, 3.5 µm (OPTIONAL)
Column Poroshell 120 EC-C18, 4.6 x 50 mm, 2.7 µm (OPTIONAL)
Column SB-C18, 4.6 x 150 mm, 5 µm (OPTIONAL)
Agilent 1260 Infinity Binary LC Optimization Guidenot orderable
Solvent Cabinet Kit (OPTIONAL)
Lab Advisor incl. license (OPTIONAL)
Items identified as "optional" are additional accessories. They are not included in the
standard scope of delivery.
Items identified as "not orderable" can be downloaded from the Agilent website
Either one of the three columns listed will be part of the delivery (as ordered).
1260 Binary Pump VL User Manual
Installing the Pump
Unpacking the Binary Pump
Accessory Kit (p/n G1311-68755)
Waste tube, 5 m (reorder pack)
Tubing assembly, i.d. 6 mm, o.d. 9 mm, 1.2 m (to waste)
CAN cable, Agilent module to module, 1 m
StS Capillary 0.17 mm, 900 mm, pump to thermostatted autosampler
StS Capillary 0.17 mm, 400 mm, pump to injector
Tubing clip (2x), re-order 4/pk
1260 Binary Pump VL User Manual
Installing the Pump
Optimizing the Stack Configuration
Optimizing the Stack Configuration
If your module is part of a complete Agilent 1260 Infinity Liquid Chromatograph, you can ensure optimum performance by installing the following configurations. These configurations optimize the system flow path, ensuring minimum delay volume.
One Stack Configuration
Ensure optimum performance by installing the modules of the Agilent 1260 Infinity LC System in the following configuration (See and . This configuration optimizes the flow path for minimum delay volume and minimizes the bench space required.
1260 Binary Pump VL User Manual
Installing the Pump
Optimizing the Stack Configuration
Column compartment
Recommended Stack Configuration for 1260 Infinity (Front View)
1260 Binary Pump VL User Manual
Installing the Pump
Optimizing the Stack Configuration
Analog detectorsignal
(1 or 2 outputsper detector)
LAN to LC ChemStation
(location depends on detector)
Recommended Stack Configuration for 1260 Infinity (Rear View)
1260 Binary Pump VL User Manual
Installing the Pump
Optimizing the Stack Configuration
Two Stack Configuration
To avoid excessive height of the stack when the autosampler thermostat is added to the system it is recommended to form two stacks. Some users prefer the lower height of this arrangement even without the autosampler thermostat. A slightly longer capillary is required between the pump and autosampler. (See
Column compartment
Degasser (optional)
Thermostat for the ALS
Recommended Two Stack Configuration for 1260 Infinity (Front View)
1260 Binary Pump VL User Manual
Installing the Pump
Optimizing the Stack Configuration
LAN to control software
(to Instant Pilot)
Recommended Two Stack Configuration for 1260 Infinity (Rear View)
1260 Binary Pump VL User Manual
Installing the Pump
Installing the Binary Pump
Installing the Binary Pump
Data Systemand/or
For other cables see text below and
Locate bench space.
Provide power connections.
Unpack the module.
Module is partially energized when switched off, as long as the power cord is
plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the
cover is opened and the module is connected to power.
➔ Make sure that it is always possible to access the power plug.
➔ Remove the power cable from the instrument before opening the cover.
➔ Do not connect the power cable to the Instrument while the covers are removed.
"Defective on arrival" problems
If there are signs of damage, please do not attempt to install the module. Inspection by Agilent is required to evaluate if the instrument is in good condition or damaged.
➔ Notify your Agilent sales and service office about the damage.
➔ An Agilent service representative will inspect the instrument at your site and
initiate appropriate actions.
1260 Binary Pump VL User Manual
Installing the Pump
Installing the Binary Pump
1 Place the module on the bench in a horizontal position.
2 Ensure the power switches on the front of the modules are OFF (switches
HiVijh>cY XVidg
Front of Binary Pump
3 At the rear of the binary pump move the security lever to its maximum right
4 Connect the power cable to the power connector at the rear of the module.
The security lever will prevent that the cover is opened while the power cord is connected to the module.
1260 Binary Pump VL User Manual
Installing the Pump
Installing the Binary Pump
5 Connect the required interface cables to the rear of the binary pump, see
Slot for interface board
Configuration Switch
Rear of the Binary Pump
6 Connect the capillary, solvent tubes and waste tubings (see
7 Press the power switch to turn on the module.
The power switch stays pressed in and a green indicator lamp in the power switch is on
when the module is turned on. When the line power switch stands out and the green light is off, the module is turned off.
8 Purge the binary pump (see
The pump was shipped with default configuration settings. To change these settings, see
1260 Binary Pump VL User Manual
Installing the Pump
Connecting Modules and Control Software
Connecting Modules and Control Software
Use of unsupplied cables
Using cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
➔ Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
Connecting Agilent 1260 Infinity Modules
1 Place the individual modules in a stack configuration as shown in
2 Ensure the power switches on the front of the modules are OFF (switches
3 Plug a CAN cable into the CAN connector at the rear of the respective
module (except vacuum degasser).
4 Connect the CAN cable to the CAN connector of the next module, see
5 Press in the power switches to turn on the modules.
1260 Binary Pump VL User Manual
Installing the Pump
Connecting Modules and Control Software
Connecting an Agilent 1260 Infinity Vacuum Degasser
1 Place the vacuum degasser in the stack of modules as shown in
2 Ensure the power switch at the front of the vacuum degasser is OFF (switch
stands out).
3 Plug an APG cable into the APG remote connector at the rear of the
4 Connect the APG cable to the APG remote connector of the pump, see
5 Press in the power switch to turn on the vacuum degasser.
The AUX output is intended for troubleshooting. It provides a DC voltage in the range of 0 –
1 V which is proportional to the vacuum level in the degasser chambers.
1260 Binary Pump VL User Manual
Installing the Pump
Connecting Modules and Control Software
Connecting Control Software and/or G4208 A Instant Pilot
With the introduction of the Agilent 1260 Infinity, all GPIB interfaces have been removed.
The preferred communication is LAN.
Usually the detector is producing the most data in the stack, followed by the pump, and it is
therefore highly recommended to use either of these modules for the LAN connection.
1 Ensure the power switches on the front of the modules in the stack are OFF
(switches stand out).
2 If there are no other 1260 with LAN port in the HPLC stack, install a
G1369B LAN board into the extension slot of the pump.
3 Connect the LAN enabled module with a LAN cable to the data system.
4 Plug the CAN connector of the Instant Pilot into any available CAN port of
the 1260 system.
5 Plug a CAN cable into the CAN connector of the Instant Pilot.
The Standard Degasser must not be connected to LAN or CAN as its connector is for
diagnostic use only.
6 Connect the CAN cable to the CAN connector of one of the modules.
7 Press in the power switches to turn on the modules.
The Agilent control software can also be connected to the system through a LAN cable,
which requires the installation of a LAN-board. For more information about connecting the Instant Pilot or Agilent control software refer to the respective user manual. des information on how to connect external hardware.
1260 Binary Pump VL User Manual
Installing the Pump
Flow Connections of the Binary Pump with Optional Solvent Selection Valve
Flow Connections of the Binary Pump with Optional Solvent
Selection Valve
wrenches 1/4 - 5/16 inch for capillary connections
Pump is installed in the LC system
When opening capillary or tube fittings, solvents may leak out.
The handling of toxic and hazardous solvents and reagents can carry health risks.
➔ Observe appropriate safety procedures (for example, wear goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.
1260 Binary Pump VL User Manual
Installing the Pump
Flow Connections of the Binary Pump with Optional Solvent Selection Valve
1 Remove the front cover by pressing the snap fasteners on both sides.
Figure 10
Removing the Front Cover
2 Place the solvent cabinet on top of the module.
3 Set the four bottles into the solvent cabinet and screw a bottle head
assembly onto each bottle.
4 Connect the solvent tubes from the bottle head assemblies to the inlet
connectors A1, A2, B1 and B2 of the solvent selection valve and label the tubes accordingly. Fix the tubes in the clips of solvent cabinet and binary pump.
5 Using a piece of sanding paper, connect the waste tubing to the purge valve
and place it into your waste system.
6 If the binary pump is not part of an Agilent 1260 Infinity system stack or
placed on the bottom of a stack, connect the corrugated waste tube to the waste outlet of the pump leak handling system.
7 Connect the outlet capillary (binary pump to injection device) to the outlet
of the purge valve.
1260 Binary Pump VL User Manual
Installing the Pump
Flow Connections of the Binary Pump with Optional Solvent Selection Valve
8 Prime your system before first use (see
Bottle-head assembly
Active seal wash option
Active inlet valve A
Active inlet valve B
Outlet capillary to autosampler
Solvent selction valve
Figure 11
Binary Pump with Solvent Selection Valve
1260 Binary Pump VL User Manual
Installing the Pump
Flow Connections of the Binary Pump Without Solvent Selection Valve
Flow Connections of the Binary Pump Without
Solvent Selection Valve
wrenches 1/4 - 5/16 inch for capillary connections
Pump is installed in the LC system
When opening capillary or tube fittings, solvents may leak out.
The handling of toxic and hazardous solvents and reagents can carry health risks.
➔ Observe appropriate safety procedures (for example, wear goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.
1260 Binary Pump VL User Manual
Installing the Pump
Flow Connections of the Binary Pump Without Solvent Selection Valve
1 Remove the front cover by pressing the snap fasteners on both sides.
Figure 12
Removing the Front Cover
2 Place the solvent cabinet on top of the binary pump.
3 Place the bottles into the solvent cabinet and place a bottle head assembly
into each bottle.
4 Connect the solvent tubes from the bottle head assemblies to the inlet
adapters of the active inlet valves. Fix the tubes in the clips of solvent cabinet and binary pump.
5 Using a piece of sanding paper, connect the waste tubing to the purge valve
and place it into your waste system.
6 If the binary pump is not part of an Agilent 1260 Infinity system stack or
placed on the bottom of a stack, connect the corrugated waste tube to the waste outlet of the pump leak handling system.
7 Connect the outlet capillary (binary pump to injection device) to the outlet
of the purge valve.
1260 Binary Pump VL User Manual
Installing the Pump
Flow Connections of the Binary Pump Without Solvent Selection Valve
8 Purge your system before first use (see
Bottle-head assembly
Pump head channel B
Pump head channel A
AdapterActive inlet valve A
Active inlet valve B
Outlet capillary to autosampler
Figure 13
Flow Connection of Binary Pump Without Solvent Selection Valve
1260 Binary Pump VL User Manual
Installing the Pump
Priming the System
Priming the System
Before a degasser or solvent tubing can be used, it is necessary to prime the system. Isopropanol is recommended as priming solvent due to its miscibility with nearly all HPLC solvents and its excellent wetting properties.
Connect all modules hydraulically as described in the respective module manuals.
Fill each solvent bottle with 100 mL isopropanol
Switch the system on
When opening capillary or tube fittings, solvents may leak out.
The handling of toxic and hazardous solvents and reagents can carry health risks.
➔ Observe appropriate safety procedures (for example, wear goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.
The purge tool of the LabAdvisor or Instrument Utilities can be used to purge the pump
If the pump is not able to draw in the solvent from the bottles, use a syringe to move the
solvent manually through tubing and degasser.
1260 Binary Pump VL User Manual
Installing the Pump
Priming the System
When priming the vacuum degasser with a syringe, the solvent is drawn through the
degasser tubes very quickly. The solvent at the degasser outlet will therefore not be fully degassed. Pump for approximately 10 minutes at your desired flow rate before starting an analysis. This will allow the vacuum degasser to properly degas the solvent in the degasser tubes.
1 Open the purge valve of the pump
2 Set the flow rate to 5 mL/min.
3 Select channel A1
4 Turn the flow on
5 Observe if the solvent in the tubing of channel A1 is advancing towards the
pump. If it isn't, disconnect the solvent tubing from the solvent selection valve, attach a syringe with a syringe adapter and pull the liquid through the degasser. Reattach the tubing to the solvent selection valve.
6 Pump 30 mL isopropanol to remove residual air bubbles.
7 Switch to the next solvent channel and repeat steps 5 and 6 until all
channels have been purged.
8 Turn the flow off and close the purge valve.
1260 Binary Pump VL User Manual
Installing the Pump
Priming the System
When the pumping system has been turned off for a certain time (for example, overnight) air will rediffuse into the solvent channel between the vacuum degasser and the pump. If solvents containing volatile components are left in the degasser without flow for a prolonged period, there will be a slight loss of the volatile components.
Switch the system on
The purge tool of the LabAdvisor or Instrument Utilities can be used for automatically
purging the pump.
1 Open the purge valve of your pump by turning it counterclockwise and set
the flow rate to 5 mL/min.
2 Flush the vacuum degasser and all tubes with at least 10 mL of solvent.
3 Repeat step 1 and 2 for the other channel(s) of the pump.
4 Set the required composition and flow rate for your application and close
the purge valve.
5 Pump for approximately 10 minutes before starting your application.
1260 Binary Pump VL User Manual
Installing the Pump
Priming the System
When the solvent of a channel is to be replaced by another solvent that is not compatible (solvents are immiscible or one solvent contains a buffer), it is necessary to follow the procedure below to prevent clogging of the pump by salt precipitation or residual liquid droplets in parts of the system.
Purging solvent(s), see
Remove the column and replace it by a ZDV fitting.
Prepare bottles with appropriat
1 If the channel is not filled with buffer, proceed to step 4.
2 Place the solvent intake filter into a bottle of water.
3 Flush the channel at a flow rate suitable for the installed tubing (typically
3 – 5 mL/min) for 10 min.
4 Modify the flow path of your system as required for your application. For
delay volume optimization, see the Rapid Resolution System manual.
Buffer salt of aqueous buffers may precipitate in residual isopropanol.
Capillaries and filter may be clogged by precipitating salt.
➔ Flush solvent lines containing high concentration of salts first with water before
introducing organic solvent.
➔ Do not perform steps 5 to 7 for channels running with aqueous buffer as solvent.
5 Replace the solvent bottle by a bottle of isopropanol.
6 Flush the channel at a flow rate suitable for the installed tubing (typically
3 – 5 mL/min) for 5 min.
7 Swap the bottle of isopropanol with a bottle of solvent for your application.
8 Repeat steps 1 to 7 for the other channel(s) of the pump.
9 Install the desired column, set the required composition and flow rate for
your application and equilibrate the system for approx. 10 minutes prior to starting a run.
1260 Binary Pump VL User Manual
Installing the Pump
Priming the System
Choice of Priming Solvents for Different Purposes
After an installation
Best solvent to flush air out of
When switching between
reverse phase and normal
Miscible with almost all
phase (both times)
After an installation
Ethanol or methanol
Alternative to isopropanol (second choice) if no isopropanol is available
To clean the system when
Best solvent to re-dissolve
After changing aqueous
Best solvent to re-dissolve
After the installation of normal
Hexane + 5 % isopropanol
Good wetting properties
phase seals (PE seals (pack of 2) (p/n 0905-1420))
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
4
Using the Binary Pump
This chapter provides information for optimized usage of the binary pump.
Using the Binary Pump
Hints for Successful Use of the Binary Pump
Hints for Successful Use of the Binary Pump
• Place solvent cabinet with the solvent bottles always on top (or at a higher
level) of the pump.
When using the binary pump without vacuum degasser, shortly degas your solvents by putting the solvent to a suitable container and applying a vacuum pressure for some time. If possible apply solvent conditions that will decrease the gas solubility over time (for example, warming up the solvents).
• The use of a vacuum degasser is mandatory for flow rates below
0.5 mL/min and for configurations without damper and mixer.
• When using the binary pump with vacuum degasser, flush the degasser with
at least 5 mL per channel before operating the pump, especially when the pumping system had been turned off for a certain length of time (for example, overnight) and volatile solvent mixtures are used in the channels (see
• Prevent blocking of solvent inlet filters (never use the pump without solvent
inlet filters). Growth of algae should be avoided (see
• Check purge valve frit and column frit in regular time intervals. A blocked
purge valve frit can be identified by black, yellow or greenish layers on its surface or by a pressure greater than 10 bar in low delay volume configuration and 20 bar in standard configuration when pumping distilled water at a rate of 5 mL/min with an open purge valve.
• Whenever possible use a minimum flow rate of 5 µL/min per solvent
channel to avoid crossflow of solvent into the unused pump channel.
• Whenever exchanging the pump seals, the purge valve frit should be
exchanged, too.
• When using buffer solutions, flush the system with water before switching it
off. The seal wash option should be used when buffer solutions with concentrations of 0.1 M or higher are being pumped for long periods of time.
• Check the pump pistons for scratches, grooves and dents when changing
the piston seals. Damaged pistons cause micro leaks and will decrease the lifetime of the seals.
• After changing the piston seals, apply the seal wear-in procedure (s
• Place the aqueous solvent on channel A and the organic solvent on
channel B. The default compressibility settings are set accordingly.
1260 Binary Pump VL User Manual
Using the Binary Pump
Setting up the Pump with the G4208A Instant Pilot
Setting up the Pump with the G4208A Instant Pilot
Generic operation of the G4208A Instant Pilot is covered in the Agilent Instant Pilot G4208A User's Guide (p/n G4208-90006). Details about setting up module specific parameters can be found in the Instant Pilot online help.
The pump parameters are described in detail in
1260 Binary Pump VL User Manual
Using the Binary Pump
Setting up the Agilent 1260 Infinity Binary Pump VL G1312C with the Instrument Control Interface
Setting up the Agilent 1260 Infinity Binary Pump VL G1312C with
the Instrument Control Interface
Parameters described in following sections is offered by the instrument control interface and can usually be accessed through Agilent instrument control software. For details, please refer to manuals and online help of respective user interfaces.
1260 Binary Pump VL User Manual
Using the Binary Pump
Setting up the Agilent 1260 Infinity Binary Pump VL G1312C with the Instrument Control Interface
Setup of Basic Pump Parameters
The most important parameters of the pump are listed in
Basic pump parameters
0.001 – 5 mL/min
Total flow rate of the pump. The optimum flow rate range is 0.1 to 5 mL/min, see .
Sto p Time
0.01 min - no limit
The stop time of the pump usually controls the run time of the whole LC system,
which is the time during which data is acquired and saved to data files. The stop
time does not stop the pump flow etc. Use no limit to stop the run manually
(useful for method development).
Po st Time
Time between the end of a run and the start of the next. Used for column equilibration after a gradient.
Pressure
Max: 0 – 400 bar
Max must be bigger than Min! Set max pressure to the maximum operating
Min: 0 – 400 bar
pressure of your column. A min pressure setting of e.g. 10 bar will turn off your pump automatically when running out of solvent. A smarter way, however, is to use the bottle fillings function (see .
Solvent A
Although channel A can be set to 0 %, it cannot be turned off. This channel should be used for the aqueous phase (water).
Solvent B
The percentage of channel B is automatically complemented by channel A to give 100 %.
max. number of lines
Use the timetable to build solvent gradients, flow gradients, or combinations of
depends on free space both. Gradients are always linear. Use multiple timetable entries to mimic in pump memory
exponential or parabolic gradients.
There are three ways to display the timetable:
•
as flow/pressure graph
as solvent percentage plot
Values can only be changed in tabular view.
1260 Binary Pump VL User Manual
Using the Binary Pump
Setting up the Agilent 1260 Infinity Binary Pump VL G1312C with the Instrument Control Interface
Pump Control
The pump can be switched between following states: On, Off or to Standby. In
Standby, the pump motor is still controlled. When the pump is switched on
from standby, it does not re-initialize.
Upon initialization, the pump ignores the Maximum Flow Gradient value.
This can result in a rapid and uncontrolled pressure increase.
➔ To prevent harm to the column, open the purge valve until the initialization is
The optional seal wash pump can be controlled by either switching it off, using it for a single time or specifying frequency and duration of periodic wash intervals.
1260 Binary Pump VL User Manual
Using the Binary Pump
Setting up the Agilent 1260 Infinity Binary Pump VL G1312C with the Instrument Control Interface
Auxiliary Pump Parameters
The auxiliary pump parameters are pre-set to fit most applications. Adjustments should only be made when required. ws the available auxiliary parameters with their default values.
Upon initialization, the pump ignores the Maximum Flow Gradient value.
This can result in a rapid and uncontrolled pressure increase.
➔ To prevent harm to the column, open the purge valve until the initialization is
Auxiliary pump parameters
Maximum Flow
0.1 – 100 mL/min²
With this parameter flow rate changes can be ramped up and down slowly to
avoid pressure shocks to the column. The default value is 100 mL/min² which
in fact turns the function off.
The volume one pump piston delivers per stroke. In general, a smaller stroke
volume results in lower pump ripple. The Auto setting adjusts the strokes
dynamically to the lowest possible value.
The strokes can be set individually for pump heads A and B.
The compressibility defines the volume change of a compressed liquid. This
parameter is used for the pump control and is required for minimizing pressure
fluctuations and optimizing flow and composition precision and accuracy. For details, please refer
1260 Binary Pump VL User Manual
Using the Binary Pump
Setting up the Agilent 1260 Infinity Binary Pump VL G1312C with the Instrument Control Interface
Data Curves
The binary pump provides the possibility to store the following operational data in the data file of the Agilent data system:
• Solvent percentage for each channel,
• pump flow,
• pressure
The pressure data curve is generated from the pressure sensor readings, while %A, %B and
flow are calculated from the method settings of the pump.
For details, please refer to the online help or manual of your instrument control software.
The pump offers a powerful feature to monitor the liquid level in the solvent bottles. With total bottle volume and initial filling volume set correctly, the pump subtracts the consumed volume continuously from the initial value and stops the pump and method/sequence execution before the system runs dry or an analysis is corrupted.
The bottle filling feature fails if multiple solvent inlets are put into one solvent bottle!
➔ In that case implement a minimum pressure limit (see ) to avoid
that the pump runs dry when solvents are empty.
he available bottle filling parameters.
1260 Binary Pump VL User Manual
Using the Binary Pump
Setting up the Agilent 1260 Infinity Binary Pump VL G1312C with the Instrument Control Interface
Bottle Filling Parameters
Tot al Volume
This is the capacity (maximum possible volume) in liter of the solvent bottle. In
combination with the actual volume, this parameter is used for calculating and displaying the relative liquid level.
After filling the solvent bottles, enter the actual volumes into these boxes. The
Actual Volume must not be larger than the Total Volume of the bottle.
default: unchecked
If this option is checked, the pump won't start a new run if the solvent level in
one or more bottles is below the minimum volume. Enter a minimum volume in liter, which considers the position of the solvent inlet and size/shape of the solvent bottle such that no air is drawn if the actual volume gets close to this limit.
Turn pump
default: unchecked
If this option is checked, the pump will turn off before air is aspirated. However,
the residual solvent volume has been calculated for 1 L solvent bottles and may be too small for large bottles or other vessels.
1260 Binary Pump VL User Manual
Using the Binary Pump
Solvent Information
Observe the following recommendations on the use of solvents.
• Follow recommendations for avoiding the growth
• Small particles can permanently block capillaries and valves. Therefore,
always filter solvents through 0.4 µm filters.
• Avoid or minimize the use of solvents that may corrode parts in the flow
path. Consider specifications for the pH range given for different materials like flow cells, valve materials etc. and recommendations in subsequent sections.
Solvent compatibility for stainless steel in standard LC systems
Stainless steel is inert against many common solvents. It is stable in the presence of acids and bases in the pH range specified for standard HPLC (pH 1 – 12.5 ). It can be corroded by acids below pH 2.3 . In general following solvents may cause corrosion and should be avoided with stainless steel:
• Solutions of alkali halides, their respective acids (for example, lithium
iodide, potassium chloride, and so on) and aequous solutions of halogenes
• High concentrations of inorganic acids like nitric acid, sulfuric acid and
organic solvents especially at higher temperatures (replace, if your chromatography method allows, by phosphoric acid or phosphate buffer which are less corrosive against stainless steel).
• Halogenated solvents or mixtures which form radicals and/or acids, for
This reaction, in which stainless steel probably acts as a catalyst, occurs quickly with dried chloroform if the drying process removes the stabilizing alcohol.
• Chromatographic grade ethers, which can contain peroxides (for example,
THF, dioxane, di-isopropylether) such ethers should be filtered through dry aluminium oxide which adsorbs the peroxides.
1260 Binary Pump VL User Manual
Using the Binary Pump
• Solutions of organic acids (acetic acid, formic acid, and so on) in organic
solvents. For example, a 1 % solution of acetic acid in methanol will attack steel.
• Solutions containing strong complexing agents (for example, EDTA,
ethylene diamine tetra-acetic acid).
• Mixtures of carbon tetrachloride with 2-propanol or THF.
1260 Binary Pump VL User Manual
Using the Binary Pump
Prevent Blocking of Solvent Filters and Algae Growth
Prevent Blocking of Solvent Filters and Algae Growth
Contaminated solvents or algae growth in the solvent bottle will reduce the lifetime of the solvent filter and will influence the performance of the module. This is especially true for aqueous solvents or phosphate buffers (pH 4 to 7). The following suggestions will prolong lifetime of the solvent filter and will maintain the performance of the module.
• Use a sterile, if possible amber, solvent bottle to slow down algae growth.
• Filter solvents through filters or membranes that remove algae.
• Exchange solvents every two days or refilter.
• If the application permits add 0.0001 – 0.001 M sodium azide to the solvent.
• Place a layer of argon on top of your solvent.
• Avoid exposure of the solvent bottle to direct sunlight.
Never use the system without solvent filter installed.
1260 Binary Pump VL User Manual
Using the Binary Pump
Algae Growth in HPLC Systems
Algae Growth in HPLC Systems
The presence of algae in HPLC systems can cause a variety of problems that may be incorrectly diagnosed as instrument or application problems. Algae grow in aqueous media, preferably in a pH range of 4-8. Their growth is accelerated by buffers, for example phosphate or acetate. Since algae grow through photosynthesis, light will also stimulate their growth. Even in distilled water small-sized algae grow after some time.
Instrumental Problems Associated With Algae
Algae deposit and grow everywhere within the HPLC system causing:
• Blocked solvent filters or deposits on inlet or outlet valves resulting in
unstable flow, composition or gradient problems or a complete failure of the pump.
• Small pore high pressure solvent filters, usually placed before the injector
to plug resulting in high system pressure.
• PTFE frits blockage leading to increased system pressure.
• Column filters to plug giving high system pressure.
• Flow cell windows of detectors to become dirty resulting in higher noise
levels (since the detector is the last module in the flow path, this problem is less common).
How to Prevent and-or Reduce the Algae Problem
• Always use freshly prepared solvents, especially use demineralized water
which was filtered through about 0.2 µm filters.
• Never leave mobile phase in the instrument for several days without flow.
• Always discard old mobile phase.
• Use the amber solvent bottle (Solvent bottle, amber (p/n 9301-1450))
supplied with the instrument for your aqueous mobile phase.
• If possible add a few mg/l sodium azide or a few percent organic solvent to
the aqueous mobile phase.
1260 Binary Pump VL User Manual
Using the Binary Pump
Algae Growth in HPLC Systems
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
This chapter gives hints on how to optimize the performance or use additional devices.
Optimizing Performance
When to Use a Vacuum Degasser
When to Use a Vacuum Degasser
The pump does not necessarily require degassing. But for the following conditions the vacuum degasser is recommended:
• if your detector is used with maximum sensitivity in the low UV wavelength
• if your application requires highest injection precision, or
• if your application requires highest retention-time reproducibility
(mandatory at flow rates below 0.5 mL/min).
Operational Hints for the Vacuum Degasser
If you are using the vacuum degasser for the first time, if the vacuum degasser was switched off for any length of time (for example, overnight), or if the vacuum degasser lines are empty, you should prime the vacuum degasser before running an analysis.
The vacuum degasser can be primed either by drawing solvent through the degasser with a syringe or by pumping with the pump.
Priming the degasser with a syringe is recommended, when:
• vacuum degasser is used for the first time, or vacuum tubes are empty, or
• changing to solvents that are immiscible with the solvent currently in the
vacuum tubes.
Priming the vacuum degasser by using the pump at high flow rate (3 – 5 mL/min) is recommended, when:
• pump was turned off for a length of time (for example, during night) and
volatile solvent mixtures are used, or
• solvents have been changed.
For more information see the Agilent 1260 Infinity Standard Degasser User Manual (p/n G1322-90012).
1260 Binary Pump VL User Manual
When to Use the Active Seal Wash Option
When to Use the Active Seal Wash Option
Concentrated buffer solutions will reduce the lifetime of the seals and pistons in your binary pump. The active seal wash option allows to maintain the seal lifetime by flushing the low pressure side of the seals with a wash solvent.
The seal wash option is strongly recommended if buffer concentrations of 0.1 M or higher are used regularly with the pump.
The active seal wash option kit can be ordered by quoting Active Seal Wash Option kit (p/n G1312-68721).
The seal wash option comprises a peristaltic pump, secondary seals, gaskets, seal holders and tubing for both pump heads. A bottle of premixed water/isopropanol (90 /10 vol%) is placed in the solvent cabinet and connected to the peristaltic pump as described in the technical note that comes with the active seal wash kit.
Always use a mixture of HPLC-grade water (90 %) and isopropanol (10 %) as wash solvent. This mixture prevents bacteria growth in the wash bottle and reduces the surface tension of the water.
In order to avoid accumulation of buffer salts or impurities, regularly replace the washing
solution using fresh solvents.
The operation of the peristaltic pump can be controlled from the data system or the Instant Pilot.
For adding a seal-wash option, please contact your local Agilent Technologies service representative.
1260 Binary Pump VL User Manual
Optimizing Performance
When to Use Alternative Seals
When to Use Alternative Seals
The standard seals for the binary pump can be used for most applications. However, normal phase applications (for example, hexane) are not compatible with the standard seals. They cause extremely high abrasion and significantly shorten seal life time.
For the use with normal phase applications special polyethylene pistons seals (yellow color, PE seals (pack of 2) (p/n 0905-1420)) are available. These seals have less abrasion compared to the standard seals.
The seal wear-in procedure causes problems to the normal phase seals (yellow).
They will be destroyed by the procedure.
➔ DO NOT apply the seal wear-in procedure performed to normal phase seals.
1 Remove the standard seals from the pump
2 Install normal phase seals.
Polyethylene seals have a limited pressure range of 0–200 bar. When used above 200 bar,
their lifetime will be significantly reduced.
1260 Binary Pump VL User Manual
When to Remove the Static Mixer
When to Remove the Static Mixer
The binary pump is equipped with a static mixer. The total delay volume of the pump is 600 – 900 µl. The mixer has a volume of 420 µl.
The static mixer and both connecting capillaries can be replaced by a small capillary (G1312-67301) under the following conditions:
• the delay volume of the pump should be reduced to a minimum for fastest
gradient response, and
• the detector is used at medium or low sensitivity.
Removing the mixer will result in an increase of the composition ripple and higher detector
1260 Binary Pump VL User Manual
Optimizing Performance
How to Optimize the Compressibility Compensation Setting
How to Optimize the Compressibility Compensation Setting
The compressibility compensation default settings are 50 × 10-6 /bar (best
for most aqueous solutions) for pump head A and 115 × 10-6 /bar (to suit
organic solvents) for pump head B. The settings represent average values
for aqueous solvents (A side) and organic solvents (B side). Therefore it is
always recommended to use the aqueous solvent on the A side of the pump
and the organic solvent on the B side. Under normal conditions the default
settings reduce the pressure pulsation to values (below 1 % of system
pressure) that will be sufficient for most applications. If the
compressibility values for the solvents used differ from the default settings,
it is recommended to change the compressibility values accordingly.
Compressibility settings can be optimized by using the values for various
solvents described in If the solvent in use is not listed in
the compressibility table, when using premixed solvents and if the default
settings are not sufficient for your application the following procedure can
be used to optimize the compressibility settings:
1 Start channel A of the binary pump with the required flow rate.
2 Before starting the optimization procedure, the flow must be stable. Use
degassed solvent only. Check the tightness of the system with the pressure test (see
3 Your pump must be connected to a data system or Instant Pilot with which
the pressure and %-ripple can be monitored, or connect an external measurement device to the analog pressure output (see
4 Start the recording device with the plot mode.
5 Starting with a compressibility setting of 10 × 10-6 /bar increase the value
in steps of 10. Re-zero the integrator as required. The compressibility compensation setting that generates the smallest pressure ripple is the optimum value for your solvent composition.
1260 Binary Pump VL User Manual
How to Optimize the Compressibility Compensation Setting
Solvent Compressibility
Carbon tetrachloride
6 hroor the B channel of
your binary pump.
1260 Binary Pump VL User Manual
Optimizing Performance
How to Optimize the Compressibility Compensation Setting
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
6
Troubleshooting and Diagnostics
Overview about the troubleshooting and diagnostic features.
Troubleshooting and Diagnostics
Overview of the Pump's Indicators and Test Functions
Overview of the Pump's Indicators and Test Functions
Status Indicators
The module is provided with two status indicators which indicate the operational state (prerun, run, and error states) of the module. The status indicators provide a quick visual check of the operation of the module.
In the event of an electronic, mechanical or hydraulic failure, the module generates an error message in the user interface. For each message, a short description of the failure, a list of probable causes of the problem, and a list of suggested actions to fix the problem are provided (see chapter Error Information).
A series of test functions are available for troubleshooting and operational verification after exchanging internal components (see Tests and Calibrations).
1260 Binary Pump VL User Manual
Troubleshooting and Diagnostics
Overview of the Pump's Indicators and Test Functions
The Pressure Test is a quick test designed to determine the pressure tightness of
the system (i.e. the high pressure flow path between pump and column). After
exchanging flow path components (e.g. pump seals or injection seal), use this
test to verify the system is pressure tight, see
Leak Test
The Leak Test is a diagnostic test designed to determine the pressure tightness
of the pump components. When a problem with the pump is suspected, use
this test to help troubleshoot the pump and its pumping performance, see
1260 Binary Pump VL User Manual
Troubleshooting and Diagnostics
Status Indicators
Two status indicators are located on the front of the module. The lower left indicates the power supply status, the upper right indicates the module status.
HiVijh>cY XVidg
Figure 14
Location of Status Indicators
Power Supply Indicator
The power supply indicator is integrated into the main power switch. When the indicator is illuminated (green) the power is ON.
1260 Binary Pump VL User Manual
Troubleshooting and Diagnostics
Module Status Indicator
The module status indicator indicates one of six possible module conditions:
• When the status indicator is OFF (and power switch light is on), the module
is in a prerun condition, and is ready to begin an analysis.
• A green status indicator, indicates the module is performing an analysis
(run mode).
• A yellow indicator indicates a not-ready condition. The module is in a
not-ready state when it is waiting for a specific condition to be reached or completed (for example, immediately after changing a set point), or while a self-test procedure is running.
• An error condition is indicated when the status indicator is red. An error
condition indicates the module has detected an internal problem which affects correct operation of the module. Usually, an error condition requires attention (e.g. leak, defective internal components). An error condition always interrupts the analysis.
If the error occurs during analysis, it is propagated within the LC system, i.e. a red LED may indicate a problem of a different module. Use the status display of your user interface for finding the root cause/module of the error.
• A blinking indicator indicates that the module is in resident mode (e.g.
during update of main firmware).
• A fast blinking indicator indicates that the module is in a low-level error
mode. In such a case try to re-boot the module or try a cold-start (see en try a firmware update (seehis does not help, a main board replacement is required.
1260 Binary Pump VL User Manual
Troubleshooting and Diagnostics
User Interfaces
Depending on the user interface, the available tests vary. Some descriptions are only available in the Service Manual.
Test Functions available vs. User Interface
Instant Pilot G4208A
Agilent Lab Advisor
Leak Test
1260 Binary Pump VL User Manual
Troubleshooting and Diagnostics
Agilent Lab Advisor Software
Agilent Lab Advisor Software
The Agilent Lab Advisor software is a standalone product that can be used with or without data system. Agilent Lab Advisor software helps to manage the lab for high quality chromatographic results and can monitor in real time a single Agilent LC or all the Agilent GCs and LCs configured on the lab intranet.
Agilent Lab Advisor software provides diagnostic capabilities for all Agilent 1200 Infinity Series modules. This includes diagnostic capabilities, calibration procedures and maintenance routines for all the maintenance routines.
The Agilent Lab Advisor software also allows users to monitor the status of their LC instruments. The Early Maintenance Feedback (EMF) feature helps to carry out preventive maintenance. In addition, users can generate a status report for each individual LC instrument. The tests and diagnostic features as provided by the Agilent Lab Advisor software may differ from the descriptions in this manual. For details refer to the Agilent Lab Advisor software help files.
The Instrument Utilities is a basic version of the Lab Advisor with limited functionality required for installation, use and maintenance. No advanced repair, troubleshooting and monitoring functionality is included.
1260 Binary Pump VL User Manual
Troubleshooting and Diagnostics
Agilent Lab Advisor Software
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
Error Information
Agilent Lab Advisor Software
This chapter describes the meaning of error messages, and provides information on probable causes and suggested actions how to recover from error conditions.
1260 Binary Pump VL User Manual
What are Error Messages?
What are Error Messages?
Error messages are displayed in the user interface when an electronic, mechanical, or hydraulic (flow path) failure occurs which requires attention before the analysis can be continued (for example, repair, or exchange of consumables is necessary). In the event of such a failure, the red status indicator at the front of the module is switched on, and an entry is written into the module logbook.
1260 Binary Pump VL User Manual
Error Information
General Error Messages
General Error Messages
General error messages are generic to all Agilent series HPLC modules and may show up on other modules as well.
Error ID: 0062
The timeout threshold was exceeded.
1 The analysis was completed successfully,
Check the logbook for the occurrence and
and the timeout function switched off the
source of a not-ready condition. Restart the
module as requested.
analysis where required.
2 A not-ready condition was present during a
Check the logbook for the occurrence and
sequence or multiple-injection run for a
source of a not-ready condition. Restart the
period longer than the timeout threshold.
analysis where required.
1260 Binary Pump VL User Manual
General Error Messages
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals. A LOW signal input on pin 4 of the remote connector generates the error message.
1 Leak detected in another module with a CAN Fix the leak in the external instrument before
connection to the system.
restarting the module.
2 Leak detected in an external instrument with Fix the leak in the external instrument before
a remote connection to the system.
restarting the module.
3 Shut-down in an external instrument with a
Check external instruments for a shut-down
remote connection to the system.
4 The degasser failed to generate sufficient
Check the vacuum degasser for an error
vacuum for solvent degassing.
condition. Refer to the Service Manual for the degasser or the 1260 pump that has the degasser built-in.
1260 Binary Pump VL User Manual
Error Information
General Error Messages
Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is started, the system expects all not-ready conditions (for example, a not-ready condition during detector balance) to switch to run conditions within one minute of starting the analysis. If a not-ready condition is still present on the remote line after one minute the error message is generated.
1 Not-ready condition in one of the
Ensure the instrument showing the not-ready
instruments connected to the remote line.
condition is installed correctly, and is set up correctly for analysis.
2 Defective remote cable.
Exchange the remote cable.
3 Defective components in the instrument
Check the instrument for defects (refer to the
showing the not-ready condition.
Lost CAN Partner
Error ID: 0071
During an analysis, the internal synchronization or communication between one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or more of the modules is no longer recognized as being connected to the system, the error message is generated.
1 CAN cable disconnected.
Ensure all the CAN cables are connected correctly.
Ensure all CAN cables are installed correctly.
2 Defective CAN cable.
Exchange the CAN cable.
3 Defective main board in another module.
Switch off the system. Restart the system, and determine which module or modules are not recognized by the system.
1260 Binary Pump VL User Manual
General Error Messages
Leak Sensor Short
Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current increases above the upper limit, the error message is generated.
1 Defective leak sensor.
Please contact your Agilent service representative.
2 Leak sensor incorrectly routed, being
Please contact your Agilent service
pinched by a metal component.
Leak Sensor Open
Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak-sensor current to change within defined limits. If the current falls outside the lower limit, the error message is generated.
1 Leak sensor not connected to the main
Please contact your Agilent service
2 Defective leak sensor.
Please contact your Agilent service representative.
3 Leak sensor incorrectly routed, being
Please contact your Agilent service
pinched by a metal component.
1260 Binary Pump VL User Manual
Error Information
General Error Messages
Compensation Sensor Open
Error ID: 0081
The ambient-compensation sensor (NTC) on the main board in the module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the main board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor increases above the upper limit, the error message is generated.
1 Defective main board.
Please contact your Agilent service representative.
Compensation Sensor Short
Error ID: 0080
The ambient-compensation sensor (NTC) on the main board in the module has failed (short circuit).
The resistance across the temperature compensation sensor (NTC) on the main board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor falls below the lower limit, the error message is generated.
1 Defective main board.
Please contact your Agilent service representative.
1260 Binary Pump VL User Manual
General Error Messages
Fan Failed
Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the main board to monitor the fan speed. If the fan speed falls below a certain limit for a certain length of time, the error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are turned off to assure that the module does not overheat inside.
1 Fan cable disconnected.
Please contact your Agilent service representative.
2 Defective fan.
Please contact your Agilent service representative.
3 Defective main board.
Please contact your Agilent service representative.
4 Improperly positioned cables or wires
Please contact your Agilent service
obstructing fan blades.
1260 Binary Pump VL User Manual
Error Information
General Error Messages
Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted temperature-compensation sensor) are used by the leak algorithm to determine whether a leak is present. When a leak occurs, the leak sensor is cooled by the solvent. This changes the resistance of the leak sensor which is sensed by the leak-sensor circuit on the main board.
1 Loose fittings.
Ensure all fittings are tight.
2 Broken capillary.
Exchange defective capillaries.
3 Loose or leaking purge valve, inlet
Ensure pump components are seated correctly. If there
valve, or outlet valve.
are still signs of a leak, exchange the appropriate seal (purge valve, inlet valve, outlet valve).
4 Defective pump seals.
Exchange the pump seals.
Open Cover
Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam is removed, the fan is switched off, and the error message is generated.
1 The top foam was removed during
Please contact your Agilent service representative.
2 Foam not activating the sensor.
Please contact your Agilent service representative.
3 Dirty or defective sensor.
Please contact your Agilent service representative.
4 Rear of the module is exposed to
Ensure that the rear of module is not directly exposed to
strong direct sunlight.
strong sunlight.
1260 Binary Pump VL User Manual
Module Error Messages
Module Error Messages
Restart Without Cover
Error ID: 2502
The module was restarted with the top cover and foam open.
The sensor on the main board detects when the top foam is in place. If the module is restarted with the foam removed, the module switches off within 30 s, and the error message is generated.
1 The module started with the top cover and
Please contact your Agilent service
foam removed.
2 Rear of the module is exposed to strong
Ensure that the rear of module is not directly
direct sunlight.
exposed to strong sunlight.
Solvent Zero Counter
Error ID: 2055, 2524
Pump firmware version A.02.32 and higher allow to set solvent bottle fillings in the data system. If the volume level in the bottle falls below the specified value the error message appears when the feature is configured accordingly.
1 Volume in bottle below specified volume.
Refill bottles and reset solvent counters.
2 Incorrect setting.
Make sure the limits are set correctly.
1260 Binary Pump VL User Manual
Error Information
Module Error Messages
Pressure Above Upper Limit
Error ID: 2014, 2500
The system pressure has exceeded the upper pressure limit.
1 Upper pressure limit set too low.
Ensure the upper pressure limit is set to a value suitable for the analysis.
2 Blockage in the flowpath (after the damper).
Check for blockage in the flowpath. The following components are particularly subject to blockage: inline filter frit, needle (autosampler), seat capillary (autosampler), sample loop (autosampler), column frits and capillaries with small internal diameters (e.g. 50 µm ID).
3 Defective damper.
Please contact your Agilent service representative.
4 Defective main board.
Please contact your Agilent service representative.
1260 Binary Pump VL User Manual
Module Error Messages
Pressure Below Lower Limit
Error ID: 2015, 2501
The system pressure has fallen below the lower pressure limit.
1 Lower pressure limit set too high.
Ensure the lower pressure limit is set to a value suitable for the analysis.
2 Air bubbles in the mobile phase.
Make sure that the degasser is in flow path and works correctly. Purge the module.
Ensure solvent inlet filters are not blocked.
Inspect the pump head, capillaries and fittings for signs of a leak.
Purge the module. Run a pressure test to determine whether the seals or other module components are defective.
4 Defective damper.
Please contact your Agilent service representative.
5 Defective main board.
Please contact your Agilent service representative.
1260 Binary Pump VL User Manual
Error Information
Module Error Messages
Pressure Signal Missing
Error ID: 2016
The pressure signal of the damper is missing.
The pressure signal of the damper must be within a specific voltage range. If the pressure signal is missing, the processor detects a voltage of approximately -120 mV across the damper connector.
1 Damper disconnected.
Please contact your Agilent service representative.
2 Defective damper.
Please contact your Agilent service representative.
Valve Failed
Error ID: 2040
Valve 0 Failed: valve A1
Valve 1 Failed: valve A2
Valve 2 Failed: valve B2
Valve 3 Failed: valve B1
One of the solvent selection valves in the module failed to switch correctly.
The processor monitors the valve voltage before and after each switching cycle. If the voltages are outside expected limits, the error message is generated.
1 Solvent selection valve disconnected.
Please contact your Agilent service representative.
2 Connection cable (inside instrument) not Please contact your Agilent service representative.
3 Connection cable (inside instrument)
Please contact your Agilent service representative.
4 Solvent selection valve defective.
Exchange the solvent selection valve.
1260 Binary Pump VL User Manual
Module Error Messages
Missing Pressure Reading
Error ID: 2054
The pressure readings read by the pump ADC (analog-digital converter) are missing.
The ADC reads the pressure signal of from the damper every 1ms. If the readings are missing for longer than 10 s, the error message is generated.
1 Damper disconnected.
Please contact your Agilent service representative.
2 Defective damper.
Please contact your Agilent service representative.
3 Defective main board.
Please contact your Agilent service representative.
Error ID: 2060
At switch-on, the pump has recognized a new pump configuration.
The pump is assigned its configuration at the factory. If the active-inlet valve and pump encoder of channel B are disconnected, and the pump is rebooted, the error message is generated. However, the pump will function as an isocratic pump in this configuration. The error message reappears after each switch-on.
1 Active-inlet valve and pump encoder of
Reconnect the active-inlet valve and pump encoder
channel B disconnected.
of channel B.
1260 Binary Pump VL User Manual
Error Information
Module Error Messages
Electronic Fuse of SSV
Error ID: 2049
Following errors can only occur, if a solvent selection valve is used. The internal number in the error message is linked either to channels A or B as shown in the following table:
Valve Fuse 0: Channels A1 and A2
Valve Fuse 1: Channels B1 and B2
One of the solvent-selection valves in the pump has drawn excessive current causing the selection-valve electronic fuse to open.
1 Defective solvent selection valve.
Restart the capillary pump. If the error message appears again, exchange the solvent selection valve.
2 Defective connection cable (front panel to
Please contact your Agilent service
main board).
3 Defective main board.
Please contact your Agilent service representative.
4 1200 Series solvent selection valve installed. Replace by 1260 solvent selection valve.
1260 Binary Pump VL User Manual
Module Error Messages
Error ID: 2044
Inlet-Valve Fuse 0: Pump channel A
Inlet-Valve Fuse 1: Pump channel B
One of the active-inlet valves in the module has drawn excessive current causing the inlet-valve electronic fuse to open.
1 Defective active inlet valve.
Restart the module. If the error message appears again, exchange the active inlet valve.
2 Defective connection cable (front panel to
Please contact your Agilent service
main board).
3 Defective main board.
Please contact your Agilent service representative.
Temperature Out of Range
Error ID: 2517
Temperature Out of Range 0: Pump channel A
Temperature Out of Range 1: Pump channel B
One of the temperature sensor readings in the motor-drive circuit are out of range.
The values supplied to the ADC by the hybrid sensors must be between 0.5 V and 4.3 V. If the values are outside this range, the error message is generated.
1 Defective main board.
Please contact your Agilent service representative.
1260 Binary Pump VL User Manual
Error Information
Module Error Messages
Temperature Limit Exceeded
Error ID: 2517
Temperature Limit Exceeded 0: Pump channel A
Temperature Limit Exceeded 1: Pump channel B
The temperature of one of the motor-drive circuits is too high.
The processor continually monitors the temperature of the drive circuits on the main board. If excessive current is being drawn for long periods, the temperature of the circuits increases. If the temperature exceeds the upper limit, the error message is generated.
1 High friction (partial mechanical blockage) in Remove the pump-head assembly. Ensure there
the pump drive assembly.
is no mechanical blockage of the pump-head assembly or pump drive assembly.
2 Partial blockage of the flowpath in front of
Ensure the outlet ball valve is not blocked.
3 Defective pump drive assembly.
Please contact your Agilent service representative.
4 Defective main board.
Please contact your Agilent service representative.
1260 Binary Pump VL User Manual
Module Error Messages
Error ID: 2041, 2042
Motor-Drive Power: Pump channel A
B: Motor-Drive Power: Pump channel B
The current drawn by the pump motor exceeded the maximum limit.
Blockages in the flow path are usually detected by the pressure sensor in the damper, which result in the pump switching off when the upper pressure limit is exceeded. If a blockage occurs before the damper, the pressure increase cannot be detected by the pressure sensor and the module will continue to pump. As pressure increases, the pump drive draws more current. When the current reaches the maximum limit, the module is switched off, and the error message is generated.
1 Flow path blockage in front of the damper.
Ensure the capillaries and frits between the pump head and damper inlet are free from blockage.
2 Blocked active inlet valve.
Exchange the active inlet valve.
3 Blocked outlet ball valve.
Exchange the outlet ball valve.
4 High friction (partial mechanical blockage) in Remove the pump-head assembly. Ensure there
the pump drive assembly.
is no mechanical blockage of the pump-head assembly or pump drive assembly.
5 Defective pump drive assembly.
Please contact your Agilent service representative.
6 Defective main board.
Please contact your Agilent service representative.
7 Restriction capillary blocked at pre-mixing
Exchange restriction capillary.
1260 Binary Pump VL User Manual
Error Information
Module Error Messages
Error ID: 2046, 2050, 2510
Encoder Missing: Pump channel A
B: Encoder Missing: Pump channel B
The optical encoder on the pump motor in the module is missing or defective.
The processor checks the presence of the pump encoder connector every 2 s. If the connector is not detected by the processor, the error message is generated.
1 Defective or disconnected pump
Please contact your Agilent service representative.
encoder connector.
2 Defective pump drive assembly.
Please contact your Agilent service representative.
Error ID: 2048, 2052
Inlet-Valve Missing: Pump channel A
B: Inlet-Valve Missing: Pump channel B
The active-inlet valve in the module is missing or defective.
The processor checks the presence of the active-inlet valve connector every 2 s. If the connector is not detected by the processor, the error message is generated.
1 Disconnected or defective cable.
Ensure the pins of the active inlet valve connector are not damaged. Ensure the connector is seated securely.
2 Disconnected or defective connection
Please contact your Agilent service representative.
cable (front panel to main board).
3 Defective active inlet valve.
Exchange the active inlet valve.
1260 Binary Pump VL User Manual
Module Error Messages
Servo Restart Failed
Error ID: 2201, 2211
Servo Restart Failed: Pump channel A
B: Servo Restart Failed: Pump channel B
The pump motor in the module was unable to move into the correct position for restarting.
When the module is switched on, the first step is to switch on the C phase of the variable reluctance motor. The rotor should move to one of the C positions. The C position is required for the servo to be able to take control of the phase sequencing with the commutator. If the rotor is unable to move, or if the C position cannot be reached, the error message is generated.
1 Disconnected or defective cable.
Please contact your Agilent service representative.
2 Blocked active inlet valve.
Exchange the active inlet valve.
3 Mechanical blockage of the module.
Remove the pump-head assembly. Ensure there is no mechanical blockage of the pump-head assembly or pump drive assembly.
4 Defective pump drive assembly.
Please contact your Agilent service representative.
5 Defective main board.
Please contact your Agilent service representative.
1260 Binary Pump VL User Manual
Error Information
Module Error Messages
Pump Head Missing
Error ID: 2202, 2212
Pump Head Missing: Pump channel A
B: Pump Head Missing: Pump channel B
The pump-head end stop in the pump was not found.
When the pump restarts, the metering drive moves forward to the mechanical end stop. Normally, the end stop is reached within 20 s, indicated by an increase in motor current. If the end point is not found within 20 s, the error message is generated.
1 Pump head not installed correctly (screws
Install the pump head correctly. Ensure nothing
not secured, or pump head not seated
(e.g. capillary) is trapped between the pump
head and body.
2 Broken piston.
Exchange the piston.
1260 Binary Pump VL User Manual
Module Error Messages
Index Limit
Error ID: 2203, 2213
Index Limit: Pump channel A
B: Index Limit: Pump channel B
The time required by the piston to reach the encoder index position was too short (pump).
During initialization, the first piston is moved to the mechanical stop. After reaching the mechanical stop, the piston reverses direction until the encoder index position is reached. If the index position is reached too fast, the error message is generated.
1 Irregular or sticking drive movement.
Remove the pump head, and examine the seals, pistons, and internal components for signs of wear, contamination or damage. Exchange components as required.
2 Defective pump drive assembly.
Please contact your Agilent service representative.
1260 Binary Pump VL User Manual
Error Information
Module Error Messages
Error ID: 2204, 2214
Index Adjustment: Pump channel A
B: Index Adjustment: Pump channel B
The encoder index position in the module is out of adjustment.
During initialization, the first piston is moved to the mechanical stop. After reaching the mechanical stop, the piston reverses direction until the encoder index position is reached. If the time to reach the index position is too long, the error message is generated.
1 Irregular or sticking drive movement.
Remove the pump head, and examine the seals, pistons, and internal components for signs of wear, contamination or damage. Exchange components as required.
2 Defective pump drive assembly.
Please contact your Agilent service representative.
Error ID: 2205, 2215, 2505
Index Missing: Pump channel A
B: Index Missing: Pump channel B
The encoder index position in the module was not found during initialization.
During initialization, the first piston is moved to the mechanical stop. After reaching the mechanical stop, the piston reverses direction until the encoder index position is reached. If the index position is not recognized within a defined time, the error message is generated.
1 Disconnected or defective encoder cable.
Please contact your Agilent service representative.
2 Defective pump drive assembly.
Please contact your Agilent service representative.
1260 Binary Pump VL User Manual
Module Error Messages
Error ID: 2206, 2216
Stroke Length: Pump channel A
B: Stroke Length: Pump channel B
The distance between the lower piston position and the upper mechanical stop is out of limits (pump).
During initialization, the module monitors the drive current. If the piston reaches the upper mechanical stop position before expected, the motor current increases as the module attempts to drive the piston beyond the mechanical stop. This current increase causes the error message to be generated.
1 Defective pump drive assembly.
Please contact your Agilent service representative.
Error ID: 2207, 2217
Initialization Failed: Pump channel A
B: Initialization Failed: Pump channel B
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the time is exceeded before initialization is complete, the error message is generated.
1 Blocked active inlet valve.
Exchange the active inlet valve.
2 Defective pump drive assembly.
Please contact your Agilent service representative.
3 Defective main board.
Please contact your Agilent service representative.
1260 Binary Pump VL User Manual
Error Information
Module Error Messages
Wait Timeout
Error ID: 2053
When running certain tests in the diagnostics mode or other special applications, the pump must wait for the pistons to reach a specific position, or must wait for a certain pressure or flow to be reached. Each action or state must be completed within the timeout period, otherwise the error message is generated.
Possible Reasons for a Wait Timeout:
· Pressure not reached.
· Pump channel A did not reach the delivery phase.
· Pump channel B did not reach the delivery phase.
· Pump channel A did not reach the take-in phase.
· Pump channel B did not reach the take-in phase.
· Solvent volume not delivered within the specified time.
1 Purge valve open.
Ensure that purge valve is closed.
2 Leak at fittings, purge valve, active inlet
Ensure pump components are seated
valve, outlet valve or piston seals.
correctly. If there are still signs of a leak, exchange the appropriate seal (purge valve, active inlet valve, outlet valve, piston seal).
Exchange defective capillaries.
3 Flow changed after starting test.
Ensure correct operating condition for the special application in use.
4 Defective pump drive assembly.
Please contact your Agilent service representative.
1260 Binary Pump VL User Manual
Module Error Messages
Electronic fuse of SSV
Error ID: 2049
The electronic fuse protecting the solvent selection valve electronics has blown.
1 Recoverable error of the SSV electronic.
Restart module, the electronic fuse can recover. If not, contact Agilent service.
2 Short cut of SSV/cable
Replace cable between board and SSV
1260 Binary Pump VL User Manual
Error Information
Module Error Messages
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
8
Test Functions and Calibration
This chapter describes the tests for the module.
Test Functions and Calibration
Pressure Test
The pressure test is a quick, built-in test designed to demonstrate the pressure-tightness of the system. The test involves monitoring the pressure profile while the binary pump runs through a predefined pumping sequence. The resulting pressure profile provides information about the pressure tightness of the system.
The test begins with the initialization of both pumpheads. After initialization, pistons A1 and B1 are both at the top of their stroke. Next, pump A begins pumping solvent with a flow rate of 510 µL/min and stroke of 100 µL. The binary pump continues to pump until a system pressure of 390 bar bar is reached.
For this test channel A is active, which is directly connected to chamber 2 in channel B, see
o test the pressure tightness of the pump use the leak test, see
When the system pressure reaches 390 bar bar, the binary pump switches off. The pressure drop from this point onwards should be no more than 2 bar/min.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Positioning the Blank Nut
To test the complete system's pressure tightness, the blank nut should be positioned at the column compartment outlet (or the outlet of the last module before the detector).
If a specific component is suspected of causing a system leak, place the blank
nut immediately before the suspected component, then run the Pressure Test
again. If the test passes, the defective component is located after the blank nut.
Confirm the diagnosis by placing the blank nut immediately after the
suspected component. The diagnosis is confirmed if the test fails.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Pressure Test
Running the Pressure Test
If problems with small leaks are suspected After maintenance of flow-path components (e.g., pump seals, injection seal) to prove pressure tightness up to 400 bar
500 mL Isopropanol
Place a bottle of LC-grade isopropanol in the solvent cabinet and connect it to channel A (or channel A2 if solvent selection valve is installed).
Make absolutely sure that all parts of the flow path that are part of the test are flushed very
thoroughly with isopropanol before starting to pressurize the system! Any trace of other solvents or the smallest air bubble inside the flow path definitely will cause the test to fail!
Running the test from the Agilent Lab Monitor & Diagnostic Software
1 Select the pressure test from the test selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by slowly opening the purge valve when the test has
describes the evaluation and interpretation of the pressure test results.
For detailed instructions refer to the Agilent Lab Monitor & Diagnostic Software.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Evaluating the Results
The sum of all leaks between the pump and the blank nut will be indicated by a pressure drop of >2 bar/minute at the plateau. Note that small leaks may cause the test to fail, but solvent may not be seen leaking from a module.
Please notice the difference between an error in the test and a failure of the test! An error
means that during the operation of the test there was an abnormal termination. If a test
failed, this means that the results of the test were not within the specified limits.
If the pressure test fails:
• Ensure all fittings between the pump and the blank nut are tight. Repeat the
pressure test.
Often, it is only a damaged blank nut itself (poorly shaped from overtightening) that causes
a failure of the test. Before investigating any other possible sources of failure make sure that the blank nut you are using is in good condition and properly tightened!
• If the test fails again, insert the blank nut at the outlet of the previous
module in the stack (eg. autosampler, port 6 of the injection valve), and repeat the pressure test. Exclude each module one by one to determine which module is leaking.
• If the pump is determined to be the source of the leak, run the leak test.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Pressure Test
Potential Causes of Pressure Test Failure
System Pressure Test failed
The test will fail, if the sum of all leaks in the system (pump, autosampler or
column compartment and connections) exceeds the test limit. After isolating
and fixing the cause of the leak, repeat the System Pressure Test to confirm the
system is pressure tight.
1 Purge valve open.
Close the purge valve.
2 Loose or leaky fittings.
Tighten the fitting or exchange the capillary.
3 Pump: Damaged pump seals or pistons.
Run the Leak Rate Test to confirm the leak.
4 Loose purge valve.
Tighten the purge valve nut (14 mm wrench).
5 Autosampler: Loose or leaky fitting.
Tighten or exchange the fitting or capillary.
6 Autosampler: Rotor seal (injection valve).
Exchange the rotor seal.
7 Autosampler: Damaged metering seal or
Exchange the metering seal. Check the piston
for scratches. Exchange the piston if required.
8 Autosampler: Needle seat.
Exchange the needle seat.
9 Column compartment: Loose or leaky fitting.
Tighten or exchange the fitting or capillary.
10 Column compartment: Rotor seal in optional Exchange the rotor seal.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Leak Test
Leak Test
The leak test is a built-in troubleshooting test designed to demonstrate the leak-tightness of the binary pump. The test involves monitoring the pressure profile as the binary pump runs through a predefined pumping sequence. The resulting pressure profile provides information about the pressure tightness and operation of the binary pump comp
The test begins with the initialization of both pumps. After initialization, pistons A1 and B1 are both at the top of their stroke. Next, the pump begins pumping solvent with a flow rate of 150 µL/min, stroke of 100 µL, and a composition of 51 % A, 49 % B. Both pumps deliver for one complete pump cycle. At the end of this step, pistons A1 and B1 are at the top of their stroke.
The pump continues pumping solvent with a flow rate of 150 µL/min. Channel A delivers for one pump cycle (first, piston A2 delivers, then piston A1), followed by channel B (piston B2, then piston B1), both channels with a stroke of 20 µL.
Just before the start of the first plateau, piston A2 delivers with a flow rate of 50 µL/min for approximately 8 s.
Plateau 1
At plateau 1, piston A2 delivers with a flow rate of 3 µL/min for 30 s.
Piston B2 delivers 50 µL/min for approximately 8 s.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Leak Test
Plateau 2
Piston B2 delivers with a flow rate of 3 µL/min for 30 s.
Piston A1 delivers 50 µL/min for approximately 8 s.
Plateau 3
Piston A1 delivers with a flow rate of 3 µL/min for 30 s.
Piston B1 delivers 50 µL/min for approximately 7 s.
Plateau 4
Piston B1 delivers with a flow rate of 3 µL/min for approximately 30 s. At the end of the fourth plateau, the test is finished and the pump switches off.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Leak Test
Running the Leak Test
If problems with the pump are suspected
Restriction Capillary
500 mL Isopropanol
Place two bottles of LC-grade isopropyl alcohol in channels A and B. If a solvent selection valve is installed, place the LC grade isopropanol in channels A2 and B2.
Make absolutely sure that all parts of the flow path that are part of the test are very
thoroughly flushed with isopropanol before starting to pressurize the system! Any trace of other solvents or the smallest air bubble inside the flow path definitely will cause the test to fail!
Running the test from the Agilent Lab Advisor
1 Select the Leak Test from the Test Selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by slowly opening the purge valve when the test has
he evaluation and interpretation of the leak test results.
Detailed instructions are provided in the Lab Advisor Software.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Leak Test
Evaluating the Results
Defective or leaky components in the pump head lead to changes in the Leak
Test pressure plot. Typical failure modes are described below.
Please notice the difference between an error in the test and a failure of the test! An error
means that during the operation of the test there was an abnormal termination. If a test failed, this means that the results of the test were not within the specified limits.
Often it is only the damaged blank nut itself (poorly shaped from overtightening) that
causes a failure of the test. Before investigating on any other possible sources of failure make sure that the blank nut you are using is in good condition and properly tightened!
No pressure increase or minimum pressure of plateau 1 not reached
1 Pump not running.
Check the logbook for error messages.
2 Purge valve open.
Close the purge valve, and restart the test.
3 Wrong solvent-line connections to solvent
Ensure the solvent lines from the degasser to
selection valve.
the solvent selection valve are connected correctly.
4 Loose or leaky fittings.
Ensure all fittings are tight, or exchange capillary.
5 Large leaks (visible) at the pump seals.
Exchange the pump seals.
6 Large leaks (visible) at active inlet valve,
Ensure the leaky components are installed
outlet valve, or purge valve.
tightly. Exchange the component if required.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Leak Test
Pressure limit not reached but plateaus horizontal or positive
1 Degasser and pump channels A and/or B
Purge the degasser and pump channels
not flushed sufficiently (air in the channels).
thoroughly with isopropanol under pressure (use the restriction capillary).
2 Wrong solvent.
Install isopropanol. Purge the degasser and pump channels thoroughly.
All plateaus negative
1 Loose or leaky fittings.
Ensure all fittings are tight, or exchange capillary.
2 Loose purge valve.
Tighten the purge valve (14 mm wrench).
3 Leaky mixer (if installed).
Tighten the mixer fittings and nuts.
4 Contaminated purge valve.
Open and close purge valve to flush out contamination. Exchange the valve if still leaky.
5 Loose pump head screws in channel A or B.
Ensure the pump head screws in channels A and B are tight.
6 Leaking seal or scratched piston in channel
Exchange the pump seals in both channels.
Check the pistons for scratches. Exchange if scratched.
7 Leaking outlet valve in channel A or B.
Exchange the outlet valve.
8 Leaky damper.
Exchange damper.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Leak Test
First plateau negative or unstable, and at least one other plateau positive
1 Leaking outlet valve in channel A.
Clean the outlet valve in channel A. Ensure the sieve in the outlet valves are installed correctly. Tighten the outlet valve.
2 Loose pump head screws in channel A.
Ensure the pump head screws in channel A are tight.
3 Leaking seal or scratched piston in channel
Exchange the pump seals in channel A. Check
the piston for scratches. Exchange if scratched.
Second plateau negative or unstable, and at least one other plateau
positive
1 Leaking outlet valve in channel B.
Clean the outlet valve in channel B. Ensure the sieve in the outlet valves are installed correctly. Tighten the outlet valve.
2 Loose pump head screws in channel B.
Ensure the pump head screws in channel B are tight.
3 Leaking seal or scratched piston in channel
Exchange the pump seals in channel B. Check
the piston for scratches. Exchange if scratched.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Leak Test
Third plateau negative or unstable and at least one other plateau positive
1 Air in channel A or new seals not yet seated. Flush channel A thoroughly with isopropanol
under pressure (use restriction capillary).
2 Loose active inlet valve in channel A.
Tighten the active inlet valve in channel A (14 mm wrench). Do not overtighten!
3 Loose pump head screws in channel A.
Ensure the pump head screws in channel A are tight.
4 Loose outlet valve in channel A.
Ensure the sieve in the outlet valve is installed correctly. Tighten the outlet valve.
5 Leaking seal or scratched piston in channel
Exchange the pump seals in channel A. Check
the pistons for scratches. Exchange if scratched.
6 Defective active inlet valve in channel A.
Exchange the active inlet valve in channel A.
Fourth plateau negative or unstable and at least one other plateau
positive
1 Air in pump chamber of channel B or seals
Flush channel B thoroughly with isopropanol
under pressure (restriction capillary).
2 Loose active inlet valve in channel B.
Tighten the active inlet valve in channel B (14mm wrench). Do not overtighten!
3 Loose pump head screws in channel B.
Ensure the pump head screws in channel B are tight.
4 Loose outlet valve in channel B.
Ensure the sieve in the outlet valve is installed correctly. Tighten the outlet valve.
5 Leaking seal or scratched piston in channel
Exchange the pump seals in channel B. Check
the pistons for scratches. Exchange if scratched.
6 Defective active inlet valve in channel B.
Exchange the active inlet valve in channel B.
1260 Binary Pump VL User Manual
Test Functions and Calibration
Leak Test
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
This chapter describes the maintenance of the module.
Maintenance
Introduction to Maintenance and Repair
Introduction to Maintenance and Repair
The module is designed for easy repair. The most frequent repairs such as piston seal change and purge valve frit change can be done from the front of the module with the module in place in the system stack.
These repairs are described in
1260 Binary Pump VL User Manual
Warnings and Cautions
Warnings and Cautions
Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
➔ When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the vendor, and follow good laboratory practice.
➔ The volume of substances should be reduced to the minimum required for the
➔ Do not operate the instrument in an explosive atmosphere.
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the
cover is opened.
➔ Do not remove the cover of the module.
➔ Only certified persons are authorized to carry out repairs inside the module.
Personal injury or damage to the product
Agilent is not responsible for any damages caused, in whole or in part, by improper
use of the products, unauthorized alterations, adjustments or modifications to the
products, failure to comply with procedures in Agilent product user guides, or use of
the products in violation of applicable laws, rules or regulations.
➔ Use your Agilent products only in the manner described in the Agilent product user
Safety standards for external equipment
➔ If you connect external equipment to the instrument, make sure that you only use
accessory units tested and approved according to the safety standards appropriate for the type of external equipment.
1260 Binary Pump VL User Manual
Overview of Maintenance
ws the main assemblies of the binary pump. The
pump heads and its parts do require normal maintenance (for example, seal exchange) and can be accessed from the front (simple repairs).
Figure 15
Overview of Maintenance Procedures
1260 Binary Pump VL User Manual
Overview of Maintenance
Solvent selection valve,
Active inlet valve, see
Purge valve, see
1260 Binary Pump VL User Manual
The procedures described in this section can be done with the binary pump in place in the system stack.
Simple Repair Procedures
If solvent filter is blocked
Gradient performance problems, intermittent
pressure fluctuations
If internally leaking
Pressure ripple unstable, run leak test for
If internally leaking
Pressure ripple unstable, run leak test for verification
If internally leaking
Solvent dripping out of waste outlet when valve
If the frit shows
A pressure drop of > 10 bar across the frit
(5 ml/min H2O with purge valve open) indicates
contamination or
If internally leaking
Error messages "Valve failed" or "Valve Fuse"
Exchanging the pump seals, see
If pump performance
Leaks at lower pump head side, unstable
indicates seal wear
retention times, pressure ripple unstable — run
leak test for verification
Exchanging pistons,
Seal life time shorter than normally
expected — check plungers while changing the
Exchanging the wash seals, see
Leaks at lower pump head side, loss of wash
indication of leaks
Error condition, indicated by red status indicator
1260 Binary Pump VL User Manual
Cleaning the Module
Cleaning the Module
To keep the module case clean, use a soft cloth slightly dampened with water, or a solution of water and mild detergent.
Liquid dripping into the electronic compartment of your module can cause shock
hazard and damage the module
➔ Do not use an excessively damp cloth during cleaning.
➔ Drain all solvent lines before opening any connections in the flow path.
1260 Binary Pump VL User Manual
Maintenance
Checking and Cleaning the Solvent Filter
Checking and Cleaning the Solvent Filter
If solvent filter is blocked
Concentrated nitric acid (35 %)
Remove solvent inlet tube from the adapter at the AIV
Small particles can permanently block the capillaries and valves of the module.
Damage of the module.
➔ Always filter solvents.
➔ Never use the module without solvent inlet filter.
If the filter is in good condition the solvent will freely drip out of the solvent tube
(hydrostatic pressure). If the solvent filter is partly blocked only very little solvent will drip out of the solvent tube.
When opening capillary or tube fittings, solvents may leak out.
The handling of toxic and hazardous solvents and reagents can carry health risks.
➔ Observe appropriate safety procedures (for example, wear goggles, safety gloves and
protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used.
Cleaning the Solvent Filter
1 Remove the blocked solvent filter from the bottle-head assembly and place
it in a beaker with concentrated nitric acid (35%) for one hour.
2 Thoroughly flush the filter with LC grade water (remove all nitric acid,
some columns can be damaged by concentrated nitric acid; check with pH indicator).
3 Reinstall the filter.
1260 Binary Pump VL User Manual
Exchanging the Purge Valve Frit or the Purge Valve
Exchanging the Purge Valve Frit or the Purge Valve
Frit – when piston seals are exchanged or when contaminated or blocked (pressure drop of > 10 bar across the frit at a flow rate of 5 mL/min of water with purge valve opened)
Purge valve – if internally leaking
Wrench open 1/4 — 5/16 inch
Wrench open 14 mm
PTFE frits (pack of 5)
Seal cap (OPTIONAL)
Switch off pump at the main power switch
Remove the front cover
Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding leakages.
1260 Binary Pump VL User Manual
Maintenance
Exchanging the Purge Valve Frit or the Purge Valve
1 Using a 1/4 inch wrench disconnect the pump outlet
2 Using the 14 mm wrench, unscrew the purge valve and
capillary from the purge valve. Disconnect the waste tube.
remove it from the purge valve holder.
Beware of leaking solvents due to hydrostatic pressure.
3 Remove the seal cap from the purge valve.
Pump outlet capillary
4 Using a pair of tweezers or a toothpick remove the frit.
5 Place a new frit into the purge valve with the orientation
of the frit as shown below (slit in frit points to the front). Reinstall the seal cap including the gold seal.
EI;:[g il i] gddkZ
Before reinstallation always check the gold seal in the
seal cap. A deformed seal cap should be exchanged.
Next Steps:
6 Insert the purge valve into the purge valve holder and orient the waste outlet nozzle downward.
7 Tighten the purge valve and reconnect outlet capillary and waste tubing.
1260 Binary Pump VL User Manual
Removing the Pump Head Assembly
Removing the Pump Head Assembly
Exchanging pump seals
Exchanging pistons
Exchanging seals of the seal wash option
Wrench open 1/4 — 5/16 inch
Hex key 3 mm12 cm long
Hex key 4 mm15 cm long T-handle
Hex driver, ¼", slitted
Switch off the pump at the main power switch
Damage of the pump drive
Starting the pump when the pump head is removed may damage the pump drive.
➔ Never start the pump when the pump head is removed.
Both pump head assemblies use the same internal components. In addition, pump head A
is fitted with the purge valve. The following procedure describes the removal and disassembly of pump head A (left). For pump head B (right) proceed in the same way and skip steps that deal with the purge valve.
1260 Binary Pump VL User Manual
Maintenance
Removing the Pump Head Assembly
1 Remove the front cover.
2 Disconnect the capillaries at the back of the purge valve holder, the pump
head adapter and the tube at the active inlet valve. Beware of leaking solvents.
3 Using a 4 mm hexagonal key stepwise loosen and remove the two pump
1260 Binary Pump VL User Manual
Maintenance of a Pump Head without Seal Wash
Maintenance of a Pump Head without Seal Wash
In case of maintenance or pump head internal leaks
Wrench open 1/4 — 5/16 inch
Hex key 3 mm12 cm long
Hex key 4 mm15 cm long T-handle
Piston seal PTFE, carbon filled, black (pack of 2), default
PE seals (pack of 2)
Switch off the pump at the main power switch
Remove the front cover to have access to the pump mechanics
Remove the pump head,
Both pump head assemblies use the same internal components. In addition, pump head A
is fitted with the purge valve. The following procedure describes the removal and disassembly of pump head A (left). For pump head B (right) proceed in the same way and skip steps that deal with the purge valve.
1260 Binary Pump VL User Manual
Maintenance
Maintenance of a Pump Head without Seal Wash
1 Place the pump head on a flat surface. Loosen the lock
2 Remove the support rings from the piston housing and lift
screw (two turns) and while holding the lower half of the
the housing away from the pistons.
assembly (piston housing) carefully pull the pump housing away from the piston housing.
3 Check the piston surface and remove any deposits or
layers. Most suitable is polishing of the piston rod with toothpaste. Replace the piston if scratched or if dents are The best way to inspect a piston is to hold it up and visible.
watch e.g. a light bulb through the piston rod. The transparent sapphire acts as a very strong magnifier and even smallest surface abnormalities become visible.
1260 Binary Pump VL User Manual
Maintenance of a Pump Head without Seal Wash
4 Using the steel side of the insert tool carefully remove the 5 Using the plastic side of the insert tool insert the new
seal from the pump housing. Remove wear retainers, if
seals into the pump head and press them firmly in
still present.
6 Place a seal wash gasket in the recess of the support ring. 7 Reassemble the pump head assembly. Note the correct
Put the seal holder on top of the gasket.
position of the pin on the support ring.
1260 Binary Pump VL User Manual
Maintenance
Maintenance of a Pump Head without Seal Wash
8 Insert the pistons and carefully press them into the seals.
Reset the seal wear counter and liquimeter in the
Agilent Lab Advisor.
9 Tighten the lock screw.
Next Steps:
10 Reinstall the pump head assembly, s
11 If a standard seal has been installed, run the seal wear-in
12 For the normal phase seal, the purge valve frit should be
1260 Binary Pump VL User Manual
Maintenance of a Pump Head with Seal Wash
Maintenance of a Pump Head with Seal Wash
In case of maintenance or pump head internal leaks
Hex key 3 mm12 cm long
Hex key 4 mm15 cm long T-handle
Screwdriver, small flat head
Gasket, seal wash (pack of 6)
Switch off the pump at the main power switch.
Remove the front cover to have access to the pump mechanics.
Both pump head assemblies use the same internal components. In addition, pump head A
is fitted with the purge valve. The following procedure describes the removal and disassembly of pump head A (left). For pump head B (right) proceed in the same way and skip steps that deal with the purge valve.
1260 Binary Pump VL User Manual
Maintenance
Maintenance of a Pump Head with Seal Wash
1 Place the pump head on a flat surface. Loosen the lock
2 Remove the seal holder and the seal wash support rings
screw (two turns) and while holding the lower half of the
from the piston housing. Remove the seal holder from the
assembly (piston housing) carefully pull the pump
support ring assembly.
housing away from the piston housing.
Seal wash support ring
3 Check the piston surface and remove any deposits or
layers. Most suitable is polishing of the piston rod with toothpaste. Replace the piston if scratched or if dents are The best way to inspect a piston is to hold it up and visible.
watch e.g. a light bulb through the piston rod. The transparent sapphire acts as a very strong magnifier and even smallest surface abnormalities become visible.
1260 Binary Pump VL User Manual
Maintenance of a Pump Head with Seal Wash
4 Using the steel side of the insert tool carefully remove the 5 Using the plastic side of the insert tool insert the new
seal from the pump housing. Remove wear retainers, if
seals into the pump head and press them firmly in
still present.
6 Using the steel side of the insert tool remove the seal
7 Using the plastic side of the insert tool press the new
wash gasket and the wash seal from the support ring. The
wash seal (spring pointing upwards) into the recess of the
removed seal will be damaged and cannot be re-used!
support ring.
Seal wash support ring
1260 Binary Pump VL User Manual
Maintenance
Maintenance of a Pump Head with Seal Wash
8 Place a seal wash gasket in the recess of the support ring. 9 Place the support rings on the piston housing (pistons not
Put the seal holder on top of the gasket.
installed) and snap the pump housing and piston housing together. Note the correct position of the pin on the support ring.
10 Insert the pistons and carefully press them into the seals. 11 Tighten the lock screw.
12 Reinstall the pump head assembly, see
1260 Binary Pump VL User Manual
Reinstalling the Pump Head Assembly
Reinstalling the Pump Head Assembly
When reassembling the pump
Hexagonal key, 3 mm
Hexagonal key, 4 mm
1 Slide the pump head assembly onto the pump drive.
2 Using a 4 mm hexagonal key tighten the pump head screws stepwise with
increasing torque.
1260 Binary Pump VL User Manual
Maintenance
Reinstalling the Pump Head Assembly
3 Reconnect the capillaries at the back of the purge valve holder, the pump
head adapter and the tube at the active inlet valve. Reconnect the active inlet valve connector.
1260 Binary Pump VL User Manual
Seal Wear-in Procedure
Seal Wear-in Procedure
This procedure is required for black PTFE seals (standard applications, p/n 5063-6589), but it will damage the yellow PE seals (normal phase applications, p/n 0905-1420).
➔ Do not run the seal wear-in procedure if PE seals are installed in the pumphead.
1 Put a bottle with 100 ml of isopropanol in the solvent cabinet and place the
solvent intake filter of the pump head you want to wear in into this bottle.
2 Screw the PEEK adapter 1/4-28 to 10-32 (p/n 0100-1847) onto the active
inlet valve and connect the inlet tube from the bottle head directly to it.
3 Connect the Restriction capillary (p/n 5022-2159) to the purge valve.
Connect its other end to a waste container.
4 Open the purge valve and purge the system for 5 min with isopropanol at a
flow rate of 2 mL/min.
5 Close the purge valve, set the flow to a value that gives a pressure of
350 bar. Pump 15 min at this pressure to wear the seals in. The pressure can be monitored on the analog output connector of the pump, with the Instant Pilot, chromatographic data system or any other controlling device connected to your pump.
6 Turn OFF the pump, slowly open the purge valve to release the pressure
from the system, disconnect the restriction capillary and reconnect the outlet capillary to the purge valve. Reconnect the intake tubing to the solvent selection valve and the connecting tube from the solvent selection valve (if installed) to the AIV.
7 Purge your system with the solvent used for your next application.
1260 Binary Pump VL User Manual
Maintenance
Exchanging the Active Inlet Valve or its Cartridge
Exchanging the Active Inlet Valve or its Cartridge
If internally leaking (backflow)
Active inlet valve body, without cartridge
Cartridge for active inlet valve 600 bar
Switch off pump at the main power switch and unplug the power cable.
Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding leakages.
1 Remove the front cover.
2 Unplug the active inlet valve cable from the connector.
3 Disconnect the solvent inlet tube at the inlet valve (beware of leaking
Binary pumps without solvent selection valve (SSV) have an adapter installed between the
solvent line and the active inlet valve (AIV). Disconnect the solvent tubes at the adapter and remove the adapter from the AIV.
1260 Binary Pump VL User Manual
Exchanging the Active Inlet Valve or its Cartridge
4 Using a 14 mm wrench loosen the active inlet valve and remove the valve
from pump head.
Active inlet valve cable
Active inlet valve
5 Using a pair of tweezers, remove the valve cartridge from the defective
active inlet valve.
1260 Binary Pump VL User Manual
Maintenance
Exchanging the Active Inlet Valve or its Cartridge
6 Push the cartridge into the new active inlet valve.
6Xi kZ caZikVakZXVgig Y Z
6Xi kZ caZikVakZ
7 Insert the valve into the pump head. Using the 14 mm wrench turn the nut
until it is hand tight.
8 Position the valve such that the solvent inlet tube connection points
towards the front.
1260 Binary Pump VL User Manual
Exchanging the Active Inlet Valve or its Cartridge
9 Tighten the nut by turning the valve in its final position using a torque
wrench (8 Nm).
10 Reconnect the Active Inlet Valve cable to the connector at the Z-panel and
the inlet tube to the valve.
11 Reinstall the front cover.
After an exchange of the valve it may be required to pump several mL of the solvent used in
the current application before the flow stabilizes at a pressure ripple as low as it used to be when the system was still working properly.
1260 Binary Pump VL User Manual
Maintenance
Exchanging the Outlet Valve
Exchanging the Outlet Valve
if leaking internally
Wrench, 1/4 - 1/5 inch
Outlet valve 1220/1260
Switch off the pump at the main power switch
1 Using a ¼ inch wrench disconnect the absorber capillary from the outlet
2 Unscrew the valve with the 14 mm wrench and remove it from the pump
3 Do not disassemble the outlet valve, as this can damage the valve.
4 Reinstall the outlet valve and tighten it using a torque wrench (12 Nm).
1260 Binary Pump VL User Manual
Exchanging the Outlet Valve
5 Reconnect the capillary at the outlet valve.
Outlet valve capillary
1260 Binary Pump VL User Manual
Maintenance
Installation of the Solvent Selection Valve Upgrade Kit
Installation of the Solvent Selection Valve Upgrade Kit
A solvent selection valve allows you to choose between four different solvents that can be used with a binary pump. The valve switches between two solvents A1 and A2 for channel A of the left pump head and two solvents B1 and B2 for channel B of the right pump head.
Applicable modules: This kit is compatible to the 1260 Infinity Binary Pumps G1312B and G1312C.
Screwdriver Pozidriv #1
Solvent Selection Valve Upgrade Kit
If required, remove solvent tubes from the inlet valves.
The figures below show a Binary Pump G1312B. The kit can be used similarly for the Binary
Pump G1312C.
1 Remove the front cover sheet by removing both screws.
2 Plug in the connector of the solvent selection valve
1260 Binary Pump VL User Manual
Installation of the Solvent Selection Valve Upgrade Kit
3 Install the solvent selection assembly by fixing both
4 Connect the outlet for solvent A (upper row) to the inlet
indicated screws.
valve of the left pump head. Connect the outlet for solvent B (lower row) to the inlet valve of the right pump head. Put the solvent bottles into the solvent cabinet. Connect the bottle heads of solvents A1 and A2 to the inlets in the upper row, see labels on valve assembly. Connect the bottle heads of solvents B1 and B2 to the inlets in the lower row, see labels on valve assembly.
For controlling the solvent selection valve, please refer to the online help or user manual of your control software.
1260 Binary Pump VL User Manual
Maintenance
Exchanging the Solvent Selection Valve
Exchanging the Solvent Selection Valve
If leaking internally ( croossflow between the ports), or if one of the channels is blocked
Screwdriver, Pozidriv #1
Solvent Selection Valve Upgrade Kit
Switch off the pump at the main power switch
1 Lift solvent bottles out of the solvent cabinet and place
2 Disconnect all tubings from the solvent selection valve.
them on the table. Disconnect the solvent tubes from the solvent selection valve and empty the tubes into the bottles. Place the bottles back into the solvent cabinet.
1260 Binary Pump VL User Manual
Exchanging the Solvent Selection Valve
3 Using a Pozidriv screwdriver #1 loosen the holding
4 Carefully pull the valve holder out and disconnect the
screws of the valve holder.
valve cable at the connector.
5 Exchange the defective solvent selection valve.
6 Tighten the screws of the valve holder.
1260 Binary Pump VL User Manual
Maintenance
Exchanging the Solvent Selection Valve
7 Reconnect all tubings to the solvent selection valve.
After an exchange of the valve it may be required to pump several mL of solvent before the
flow stabilizes at a pressure ripple as low as it used to be when the system was still working properly.
1260 Binary Pump VL User Manual
Exchanging the Optional Interface Board
Exchanging the Optional Interface Board
BCD (Interface) board
Switch OFF the module at the main power switch.
Unplug the module from main power.
Electronic boards and components are sensitive to electrostatic discharge (ESD).
ESD can damage electronic boards and components.
➔ In order to prevent damage always use an ESD protection when handling electronic
boards and components.
1 Disconnect cables from the interface board connectors.
2 Loosen the screws. Slide out the interface board from the module.
BCD (interface) board
Figure 16
Exchanging the Interface Board
3 Install the new interface board. Secure the screws.
4 Reconnect the cables to the board connector
1260 Binary Pump VL User Manual
Maintenance
Replacing the Module Firmware
Replacing the Module Firmware
The installation of newer firmware might be necessary
•
if a newer version solves problems of older versions or
to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
•
to keep all systems on the same (validated) revision or
if a new module with newer firmware is added to a system or
if third part control software requires a special version.
LAN/RS-232 Firmware Update Tool
Agilent Diagnostic Software
Instant Pilot G4208A(only if supported by module)
Firmware, tools and documentation from Agilent web site
Read update documentation provided with the Firmware Update Tool.
To upgrade/downgrade the module's firmware carry out the following
steps:
1 Download the required module firmware, the latest LAN/RS-232 FW
Update Tool and the documentation from the Agilent web.
2 For loading the firmware into the module follow the instructions in the
Module Specific Information
There is no specific information for this module.
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
10
Parts for Maintenance
This chapter provides information on parts for maintenance.
10 Parts for Maintenance
Pump Head Assembly Without Seal Wash
Pump Head Assembly Without Seal Wash
Support Ring SL, no seal wash
Gasket, seal wash (pack of 6)
Capillary, outlet valve to piston 2
Piston seal PTFE, carbon filled, black (pack of 2), default
Pump chamber housing
Mounting screw for manual purge valve holder, M4, 20 mm long
Holder for manual purge valve
Outlet valve 1220/1260
Active inlet valve body, without cartridge
Cartridge for active inlet valve 600 bar
Pump head screw (M5, 60 mm)
1260 Binary Pump VL User Manual
Parts for Maintenance
Pump Head Assembly Without Seal Wash
Figure 17
Pump Head Assembly Without Seal Wash
1260 Binary Pump VL User Manual
10 Parts for Maintenance
Pump-Head Assembly with Seal Wash
Pump-Head Assembly with Seal Wash
Seal wash pump assembly
Peristaltic pump cartridge, silicone tubing
Support ring seal wash
Tubing (seal wash)
Gasket, seal wash (pack of 6)
Capillary, outlet valve to piston 2
Piston seal PTFE, carbon filled, black (pack of 2), default
Pump chamber housing
Mounting screw for manual purge valve holder, M4, 20 mm long
Holder for manual purge valve
Outlet valve 1220/1260
Active inlet valve body, without cartridge
Cartridge for active inlet valve 600 bar
Pump head screw (M5, 60 mm)
1260 Binary Pump VL User Manual
Parts for Maintenance
Pump-Head Assembly with Seal Wash
Figure 18
Pump head assembly with seal wash option
1260 Binary Pump VL User Manual
10 Parts for Maintenance
Outlet Valve
Outlet Valve
Outlet valve 1220/1260
Figure 19
1260 Binary Pump VL User Manual
Parts for Maintenance
Purge Valve Assembly
Purge Valve Assembly
PTFE frits (pack of 5)
Figure 20
Purge Valve Assembly
1260 Binary Pump VL User Manual
10 Parts for Maintenance
Active Inlet Valve Assembly
Active Inlet Valve Assembly
Active inlet valve body, without cartridge
Cartridge for active inlet valve 600 bar
Figure 21
Active Inlet Valve Assembly
1260 Binary Pump VL User Manual
Parts for Maintenance
Accessory Kit (p/n G1311-68755)
Waste tube, 5 m (reorder pack)
Tubing assembly, i.d. 6 mm, o.d. 9 mm, 1.2 m (to waste)
CAN cable, Agilent module to module, 1 m
StS Capillary 0.17 mm, 900 mm, pump to thermostatted autosampler
StS Capillary 0.17 mm, 400 mm, pump to injector
Tubing clip (2x), re-order 4/pk
Active Seal Wash Option
Active Seal Wash Option kit (p/n G1312-68721)
Seal wash pump assembly
Peristaltic pump cartridge, silicone tubing
Secondary seal (pre-installed in support rings)
Gasket, seal wash (pack of 6)
Silicone tubing, 1 mm i.d., 3 mm o.d., 5 m, re-order number
Standard seals (pack of 2)
01018-23702 Insert
1260 Binary Pump VL User Manual
10 Parts for Maintenance
Solvent Cabinet Kit
Front panel, solvent cabinet
Leak pan, solvent cabinet
Solvent bottle, transparent
Solvent bottle, amber
Bottle-head assembly
Figure 22
Solvent Cabinet Parts
1260 Binary Pump VL User Manual
Parts for Maintenance
Bottle Head Assembly
Bottle Head Assembly
Solvent bottle, amber
Solvent bottle, transparent
Bottle-head assembly
Ferrules with lock ring (10/Pk)
Tube screw (10/Pk)
Solvent tubing, 5 m
Inlet filter adapter (4/Pk)
Solvent inlet filter, 20 µm pore size
Figure 23
Bottle-Head Assembly Parts
1260 Binary Pump VL User Manual
10 Parts for Maintenance
Hydraulic Path with Solvent Selection Valve
Hydraulic Path with Solvent Selection Valve
Capillary, damper to mixer
Capillary, mixer to purge valve
Capillary, outlet valve to piston 2
Capillary, mixing chamber to damper inlet
Capillary, channel A and B pump head outlet to mixing chamber (included)
Waste tube, 5 m (reorder pack)
Outlet capillary, pump to injector
Capillary, pump to thermostattable autosampler
Capillary, damper to purge valveCapillary needed when pump used without mixer installed
Bottle-head assembly
1260 Binary Pump VL User Manual
Parts for Maintenance
Hydraulic Path with Solvent Selection Valve
Figure 24
Hydraulic path with solvent selection valve and active seal wash option
1260 Binary Pump VL User Manual
10 Parts for Maintenance
Hydraulic Path without Solvent Selection Valve
Hydraulic Path without Solvent Selection Valve
Bottle-head assembly
Capillary, damper to mixer
Capillary, mixer to purge valve
Capillary, outlet valve to piston 2
Capillary, mixing chamber to damper inlet
Capillary, channel A and B pump head outlet to mixing chamber (included)
PEEK adapter 1/4-28 to 10-32(Adapter AIV to solvent inlet tubes)
Waste tube, 5 m (reorder pack)
Outlet capillary, pump to injector
Capillary, pump to thermostattable autosampler
Capillary, damper to purge valveCapillary needed when pump used without mixer installed
1260 Binary Pump VL User Manual
Parts for Maintenance
Hydraulic Path without Solvent Selection Valve
Figure 25
Hydraulic Path without Solvent Selection Valve
1260 Binary Pump VL User Manual
10 Parts for Maintenance
Hydraulic Path without Solvent Selection Valve
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
This chapter provides information on cables used with the Agilent 1200 Infinity Series modules.
11 Identifying Cables
Never use cables other than the ones supplied by Agilent Technologies to ensure proper
functionality and compliance with safety or EMC regulations.
Agilent module to 3394/6 integrators
Agilent 35900A A/D converter
Analog cable (BNC to general purpose, spade lugs)
Agilent module to 3396A Series I integrators
3396 Series II / 3395A integrator, see details in section
Agilent module to 3396 Series III / 3395B integrators
Agilent module to general purpose
BCD cables
Agilent module to 3396 integrators
Agilent module to general purpose
1260 Binary Pump VL User Manual
CAN cables
CAN cable, Agilent module to module, 0.5 m
CAN cable, Agilent module to module, 1 m
LAN cables
Cross-over network cable, shielded, 3 m (for point to point connection)
Twisted pair network cable, shielded, 7 m (for point to point connection)
External Contact Cable
External contact cable - Agilent module interface board to general purposes
RS-232 cable, 2 m
RS-232 cable, 2.5 mInstrument to PC, 9-to-9 pin (female). This cable has special pin-out, and is not compatible with connecting printers and plotters. It's also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
RS-232 cable, 8 m
1260 Binary Pump VL User Manual
11 Identifying Cables
One end of these cables provides a BNC connector to be connected to Agilent modules. The other end depends on the instrument to which connection is being made.
Agilent Module to 3394/6 Integrators
Pin 3394/6
Pin Agilent
Signal Name
Agilent Module to BNC Connector
Pin Agilent
Signal Name
1260 Binary Pump VL User Manual
Agilent Module to General Purpose
Pin Agilent
Signal Name
1260 Binary Pump VL User Manual
11 Identifying Cables
One end of these cables provides a Agilent Technologies APG (Analytical Products Group) remote connector to be connected to Agilent modules. The other end depends on the instrument to be connected to.
Agilent Module to 3396A Integrators
Pin 3396A
Pin Agilent
Signal Name
Agilent Module to 3396 Series II / 3395A Integrators
Use the cable Agilent module to 3396A Series I integrators (p/n 03394-60600) and cut pin #5 on the integrator side. Otherwise the integrator prints START; not ready.
1260 Binary Pump VL User Manual
Agilent Module to 3396 Series III / 3395B Integrators
Pin Agilent
Signal Name
Agilent Module to Agilent 35900 A/D Converters
Pin 35900 A/D
Pin Agilent
Signal Name
1260 Binary Pump VL User Manual
11 Identifying Cables
Agilent Module to General Purpose
Wire Color
Pin Agilent
Signal Name
1260 Binary Pump VL User Manual
BCD Cables
BCD Cables
One end of these cables provides a 15-pin BCD connector to be connected to the Agilent modules. The other end depends on the instrument to be connected to
Agilent Module to General Purpose
Wire Color
Pin Agilent
Signal Name
BCD Digit
1260 Binary Pump VL User Manual
11 Identifying Cables
BCD Cables
Agilent Module to 3396 Integrators
Pin Agilent
Signal Name
BCD Digit
1260 Binary Pump VL User Manual
CAN Cable
CAN Cable
Both ends of this cable provide a modular plug to be connected to Agilent modules CAN or LAN connectors.
CAN Cables
CAN cable, Agilent module to module, 0.5 m
CAN cable, Agilent module to module, 1 m
LAN Cables
Cross-over network cable, shielded, 3 m (for point to point connection)
Twisted pair network cable, shielded, 7 m (for point to point connection)
1260 Binary Pump VL User Manual
11 Identifying Cables
External Contact Cable
External Contact Cable
One end of this cable provides a 15-pin plug to be connected to Agilent modules interface board. The other end is for general purpose.
Agilent Module Interface Board to general purposes
Pin Agilent
Signal Name
1260 Binary Pump VL User Manual
Agilent Module to PC
Agilent Module to PC
RS-232 cable, 2 m
RS-232 cable, 2.5 mInstrument to PC, 9-to-9 pin (female). This cable has special pin-out, and is not compatible with connecting printers and plotters. It's also called "Null Modem Cable" with full handshaking where the wiring is made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
RS-232 cable, 8 m
1260 Binary Pump VL User Manual
11 Identifying Cables
Agilent 1200 Module to Printer
Agilent 1200 Module to Printer
Cable Printer Serial & Parallel, is a SUB-D 9 pin female vs. Centronics connector on the other end (NOT FOR FW UPDATE). For use with G1323 Control Module.
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
This chapter describes the pump in more detail on hardware and electronics.
12 Hardware Information
The firmware of the instrument consists of two independent sections:
• a non-instrument specific section, called resident system
• an instrument specific section, called main system
This resident section of the firmware is identical for all Agilent 1100/1200/1220/1260/1290 series modules. Its properties are:
• the complete communication capabilities (CAN, LAN and RS-232C)
• memory management
• ability to update the firmware of the 'main system'
Main System
Its properties are:
• the complete communication capabilities (CAN, LAN and RS-232C)
• memory management
• ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are divided into common functions like
• run synchronization through APG remote,
• error handling,
• diagnostic functions,
• or module specific functions like
• internal events such as lamp control, filter movements,
• raw data collection and conversion to absorbance.
1260 Binary Pump VL User Manual
Firmware updates can be done using your user interface:
• PC and Firmware Update Tool with local files on the hard disk
• Instant Pilot (G4208A) with files from a USB Flash Disk
• Agilent Lab Advisor software B.01.03 and above
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
PPPP is the product number, for example, 1315AB for the G1315A/B DAD,
R the firmware revision, for example, A for G1315B or B for the G1315C DAD,
VVV is the revision number, for example 102 is revision 1.02,
XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware in chapter "Maintenance" or use the documentation provided with the Firmware Update Tools.
Update of main system can be done in the resident system only. Update of the resident
system can be done in the main system only.
Main and resident firmware must be from the same set.
GZh YZci;LJeYViZ
Figure 26
Firmware Update Mechanism
1260 Binary Pump VL User Manual
12 Hardware Information
Some modules are limited in downgrading due to their main board version or their initial
firmware revision. For example, a G1315C DAD SL cannot be downgraded below firmware revision B.01.02 or to a A.xx.xx.
Some modules can be re-branded (e.g. G1314C to G1314B) to allow operation in specific control software environments. In this case the feature set of the target type are use and the feature set of the original are lost. After re-branding (e.g. from G1314B to G1314C), the original feature set is available again.
All these specific informations are described in the documentation provided with the firmware update tools.
The firmware update tools, firmware and documentation are available from the Agilent web.
1260 Binary Pump VL User Manual
• The CAN bus is a serial bus with high speed data transfer. The two
connectors for the CAN bus are used for internal module data transfer and synchronization.
• One analog output provides signals for integrators or data handling
• The interface board slot is used for external contacts and BCD bottle
number output or LAN connections.
• The REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as start, stop, common shut down, prepare, and so on.
• With the appropriate software, the RS-232C connector may be used to
control the module from a computer through a RS-232C connection. This connector is activated and can be configured with the configuration switch.
• The power input socket accepts a line voltage of 100 – 240 VAC ± 10 % with a
line frequency of 50 or 60 Hz. Maximum power consumption varies by module. There is no voltage selector on your module because the power supply has wide-ranging capability. There are no externally accessible fuses, because automatic electronic fuses are implemented in the power supply.
Never use cables other than the ones supplied by Agilent Technologies to ensure proper
functionality and compliance with safety or EMC regulations.
1260 Binary Pump VL User Manual
12 Hardware Information
Rear View of the Module
Configuration switch
Slot for interface board
Figure 27
Electrical connections to the binary pump
The GPIB interface has been removed with the introduction of the 1260 Infinity modules.
1260 Binary Pump VL User Manual
The Agilent 1200 Infinity Series modules provide the following interfaces:
Agilent 1200 Infinity Series Interfaces
G1311B Quat PumpG1311C Quat Pump VLG1312B Bin PumpG1312C Bin Pump VL1376A Cap PumpG2226A Nano PumpG5611A Bio-inert Quat Pump
G4220A/B Bin Pump
CAN-DC- OUT for CAN slaves
CAN-DC- OUT for CAN
G1377A HiP micro ALSG2258A DL ALSG5664A Bio-inert FC-ASG5667A Bio-inert Autosampler
1260 Binary Pump VL User Manual
12 Hardware Information
Agilent 1200 Infinity Series Interfaces
G1365C MWDG1315D DAD VLG1365D MWD VL
EXT ContactAUTOZERO
G1170A Valve Drive
Requires a HOST module with on-board LAN (e.g. G4212A or G4220A with minimum firmware B.06.40 or C.06.40) or with additional G1369C LAN Card
CAN-DC- OUT for CAN slavesTHERMOSTAT for G1330A/B (NOT USED)
1260 Binary Pump VL User Manual
The detector (DAD/MWD/FLD/VWD/RID) is the preferred access point for control via
LAN. The inter-module communication is done via CAN.
• CAN connectors as interface to other modules
• LAN connector as interface to the control software
• RS-232C as interface to a computer
• REMOTE connector as interface to other Agilent products
• Analog output connector(s) for signal output
1260 Binary Pump VL User Manual
12 Hardware Information
Overview Interfaces
The CAN is inter-module communication interface. It is a 2-wire serial bus system supporting high speed data communication and real-time requirement.
The modules have either an interface slot for an LAN card (e.g. Agilent G1369B/C LAN Interface) or they have an on-board LAN interface (e.g. detectors G1315C/D DAD and G1365C/D MWD). This interface allows the control of the module/system via a PC with the appropriate control software.
If an Agilent detector (DAD/MWD/FLD/VWD/RID) is in the system, the LAN should be
connected to the DAD/MWD/FLD/VWD/RID (due to higher data load). If no Agilent detector is part of the system, the LAN interface should be installed in the pump or autosampler.
The RS-232C connector is used to control the module from a computer through RS-232C connection, using the appropriate software. This connector can be configured with the configuration switch module at the rear of the module. Refer to Communication Settings for RS-232C.
There is no configuration possible on main boards with on-board LAN. These are
pre-configured for
• 19200 baud,
• 8 data bit with no parity and
• one start bit and one stop bit are always used (not selectable).
The RS-232C is designed as DCE (data communication equipment) with a 9-pin male SUB-D type connector. The pins are defined as:
1260 Binary Pump VL User Manual
RS-232C Connection Table
Figure 28
Analog Signal Output
The analog signal output can be distributed to a recording device. For details refer to the description of the module's main board.
1260 Binary Pump VL User Manual
12 Hardware Information
APG Remote
The APG Remote connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features as common shut down, prepare, and so on.
Remote control allows easy connection between single instruments or systems to ensure coordinated analysis with simple coupling requirements.
The subminiature D connector is used. The module provides one remote connector which is inputs/outputs (wired- or technique).
To provide maximum safety within a distributed analysis system, one line is
dedicated to SHUT DOWN the system's critical parts in case any module detects
a serious problem. To detect whether all participating modules are switched
on or properly powered, one line is defined to summarize the POWER ON state
of all connected modules. Control of analysis is maintained by signal readiness
READY for next analysis, followed by START of run and optional STOP of run
triggered on the respective lines. In addition PREPARE and START REQUEST may
be issued. The signal levels are defined as:
• standard TTL levels (0 V is logic true, + 5.0 V is false),
• fan-out is 10 ,
• input load is 2.2 kOhm against + 5.0 V, and
• output are open collector type, inputs/outputs (wired- or technique).
All common TTL circuits operate with a 5 V power supply. A TTL signal is defined as "low"
or L when between 0 V and 0.8 V and "high" or H when between 2.0 V and 5.0 V (with respect to the ground terminal).
1260 Binary Pump VL User Manual
Remote Signal Distribution
(L) Request to prepare for analysis (for example, calibration, detector lamp on). Receiver is any module performing pre-analysis activities.
(L) Request to start run / timetable. Receiver is any module performing run-time controlled activities.
(L) System has serious problem (for example, leak: stops pump). Receiver is any module capable to reduce safety risk.
(H) All modules connected to system are switched on. Receiver is any module relying on operation of others.
(H) System is ready for next analysis. Receiver is any sequence controller.
(L) Request to reach system ready state as soon as possible (for example, stop run, abort or finish and stop injection). Receiver is any module performing run-time controlled activities.
(L) Request to start injection cycle (for example, by start key on any module). Receiver is the autosampler.
Some modules have module specific interfaces/connectors. They are described in the module documentation.
1260 Binary Pump VL User Manual
12 Hardware Information
Setting the 8-bit Configuration Switch (without On-board) LAN
Setting the 8-bit Configuration Switch (without On-board) LAN
The 8-bit configuration switch is located at the rear of the module.
This module does not have its own on-board LAN interface. It can be controlled through the LAN interface of another module, and a CAN connection to that module.
Figure 29
Configuration switch (settings depend on configured mode)
All modules without on-board LAN:
• default should be ALL DIPS DOWN (= best settings)
• Bootp mode for LAN and
• 19200 baud, 8 data bit / 1 stop bit with no parity for RS-232
• DIP 1 DOWN and DIP 2 UP allows special RS-232 settings
• for boot/test modes DIPS 1+2 must be UP plus required mode
For normal operation use the default (best) settings.
Switch settings provide configuration parameters for serial communication protocol and instrument specific initialization procedures.
With the introduction of the Agilent 1260 Infinity, all GPIB interfaces have been removed.
The preferred communication is LAN.
1260 Binary Pump VL User Manual
Setting the 8-bit Configuration Switch (without On-board) LAN
The following tables represent the configuration switch settings for the modules without
on-board LAN only.
8-bit Configuration Switch (without on-board LAN)
The LAN settings are done on the LAN Interface Card G1369B/C. Refer to the
documentation provided with the card.
Communication Settings for RS-232C
The communication protocol used in the column compartment supports only hardware handshake (CTS/RTR).
Switches 1 in down and 2 in up position define that the RS-232C parameters will be changed. Once the change has been completed, the column instrument must be powered up again in order to store the values in the non-volatile memory.
Communication Settings for RS-232C Communication (without on-board LAN)
Use the following tables for selecting the setting which you want to use for RS-232C communication. The number 0 means that the switch is down and 1 means that the switch is up.
1260 Binary Pump VL User Manual
12 Hardware Information
Setting the 8-bit Configuration Switch (without On-board) LAN
Baudrate Settings (without on-board LAN)
Baud Rate
Baud Rate
Data Bit Settings (without on-board LAN)
7 Bit Communication
8 Bit Communication
Parity Settings (without on-board LAN)
One start bit and one stop bit are always used (not selectable).
Per default, the module will turn into 19200 baud, 8 data bit with no parity.
1260 Binary Pump VL User Manual
Setting the 8-bit Configuration Switch (without On-board) LAN
The special settings are required for specific actions (normally in a service case).
Firmware update procedures may require this mode in case of firmware loading errors (main firmware part).
If you use the following switch settings and power the instrument up again, the instrument firmware stays in the resident mode. It is not operable as a module. It only uses basic functions of the operating system for example, for communication. In this mode the main firmware can be loaded (using update utilities).
Boot Resident Settings (without on-board LAN)
Mode Select
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with default parameter settings.
Forced cold start erases all methods and data stored in the non-volatile memory. Exceptions are calibration settings, diagnosis and repair log books which will not be erased.
➔ Save your methods and data before executing a forced cold start.
If you use the following switch settings and power the instrument up again, a forced cold start has been completed.
Forced Cold Start Settings (without on-board LAN)
Mode Select
1260 Binary Pump VL User Manual
12 Hardware Information
Setting the 8-bit Configuration Switch (without On-board) LAN
1260 Binary Pump VL User Manual
1260 Binary Pump VL User Manual
This chapter provides addition information on safety, legal and web.
13 Appendix
General Safety Information
General Safety Information
General Safety Information
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer's failure to comply with these requirements.
Ensure the proper usage of the equipment.
The protection provided by the equipment may be impaired.
➔ The operator of this instrument is advised to use the equipment in a manner as
specified in this manual.
This is a Safety Class I instrument (provided with terminal for protective earthing) and has been manufactured and tested according to international safety standards.
Before applying power, comply with the installation section. Additionally the following must be observed.
Do not remove instrument covers when operating. Before the instrument is switched on, all protective earth terminals, extension cords, auto-transformers, and devices connected to it must be connected to a protective earth via a ground socket. Any interruption of the protective earth grounding will cause a potential shock hazard that could result in serious personal injury. Whenever it is likely that the protection has been impaired,
1260 Binary Pump VL User Manual
General Safety Information
the instrument must be made inoperative and be secured against any intended operation.
Make sure that only fuses with the required rated current and of the specified type (normal blow, time delay, and so on) are used for replacement. The use of repaired fuses and the short-circuiting of fuse holders must be avoided.
Some adjustments described in the manual, are made with power supplied to the instrument, and protective covers removed. Energy available at many points may, if contacted, result in personal injury.
Any adjustment, maintenance, and repair of the opened instrument under voltage should be avoided whenever possible. When inevitable, this has to be carried out by a skilled person who is aware of the hazard involved. Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present. Do not replace components with power cable connected.
Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
Do not install substitute parts or make any unauthorized modification to the instrument.
Capacitors inside the instrument may still be charged, even though the instrument has been disconnected from its source of supply. Dangerous voltages, capable of causing serious personal injury, are present in this instrument. Use extreme caution when handling, testing and adjusting.
When working with solvents, observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet by the solvent vendor, especially when toxic or hazardous solvents are used.
1260 Binary Pump VL User Manual
13 Appendix
General Safety Information
The apparatus is marked with this symbol when the user should refer to the instruction manual in order to protect risk of harm to the operator and to protect the apparatus against damage.
Indicates dangerous voltages.
Indicates a protected ground terminal.
Indicates eye damage may result from directly viewing the light produced by the deuterium lamp used in this product.
The apparatus is marked with this symbol when hot surfaces are available and the user should not touch it when heated up.
A WARNING
alerts you to situations that could cause physical injury or death.
➔ Do not proceed beyond a warning until you have fully understood and met the
indicated conditions.
alerts you to situations that could cause loss of data, or damage of equipment.
➔ Do not proceed beyond a caution until you have fully understood and met the
1260 Binary Pump VL User Manual
The Waste Electrical and Electronic Equipment Directive
The Waste Electrical and Electronic Equipment Directive
The Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC), adopted by EU Commission on 13 February 2003, is introducing producer responsibility on all electric and electronic appliances starting with 13 August 2005.
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The
affixed label indicates that you must not discard this electrical/electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a Monitoring and Control Instrumentation product.
Do not dispose off in domestic household waste
To return unwanted products, contact your local Agilent office, or see www.agilent.com for more information.
1260 Binary Pump VL User Manual
13 Appendix
Lithium batteries may not be disposed-off into the domestic waste. Transportation of
discharged Lithium batteries through carriers regulated by IATA/ICAO, ADR, RID,
IMDG is not allowed.
Danger of explosion if battery is incorrectly replaced.
➔ Discharged Lithium batteries shall be disposed off locally according to national
waste disposal regulations for batteries.
➔ Replace only with the same or equivalent type recommended by the equipment
Lithiumbatteri - Eksplosionsfare ved fejlagtig håndtering.
Udskiftning må kun ske med batteri af samme fabrikat og type.
➔ Lever det brugte batteri tilbage til leverandøren.
Lithiumbatteri - Eksplosionsfare.
Ved udskiftning benyttes kun batteri som anbefalt av apparatfabrikanten.
➔ Brukt batteri returneres appararleverandoren.
Bij dit apparaat zijn batterijen geleverd. Wanneer deze leeg zijn, moet u ze niet weggooien
maar inleveren als KCA.
1260 Binary Pump VL User Manual
Cables supplied by Agilent Technologies are screened to provide optimized protection against radio interference. All cables are in compliance with safety or EMC regulations.
Test and Measurement
If test and measurement equipment is operated with unscreened cables, or used for measurements on open set-ups, the user has to assure that under operating conditions the radio interference limits are still met within the premises.
1260 Binary Pump VL User Manual
13 Appendix
This statement is provided to comply with the requirements of the German Sound Emission Directive of 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB.
• Sound Pressure Lp < 70 dB (A)
• At Operator Position
• Normal Operation
• According to ISO 7779:1988/EN 27779/1991 (Type Test)
1260 Binary Pump VL User Manual
Agilent Technologies on Internet
Agilent Technologies on Internet
For the latest information on products and services visit our worldwide web site on the Internet at:
Select Products/Chemical Analysis
It will provide also the latest firmware of the modules for download.
1260 Binary Pump VL User Manual
changing solvents
8-bit configuration switch
without On-Board LAN
Communication settings
bottle head assembly
active inlet valve
buffer application
compensation sensor open
compensation sensor short
composition precision
Agilent Diagnostic software
composition range
Agilent Lab Advisor software
Agilent Lab Advisor
connecting APG remote
alternative seal material
connecting the ChemStation
connections, flow
ambient non-operating temperature
connecting the power
ambient operating temperature
defect on arrival
Diagnostic software
dual-piston in-series design
electrical connections
safety information
1260 Binary Pump VL User Manual
active inlet valve
electrostatic discharge (ESD)
highest injection precision
outlet valve sieve
high-pressure mixing
early maintenance feedback
hints for successful use
hydraulic path with solvent selection
solvent selection valve
hydraulic path without solvent selection
compensation sensor open
compensation sensor short
ignition without cover
fastest gradient response
initialization failed
initialization failed
inlet-valve missing
inlet-valve missing
installation, pump module
leak sensor short
site requirements
missing pressure reading
instrument layout
motor-drive power
pressure above upper limit
pressure below lower limit
pressure signal missing
pump head missing
selection-valve fuse
introduction to the pump
servo restart failed
solvent zero counter
temperature limit exceeded
temperature out of range
leak sensor short
leak test, evaluation
general error messages
gradient formation
active inlet valve
1260 Binary Pump VL User Manual
active inlet valve
purge-valve assembly
bottle head assembly
lithium batteries
hydraulic path with solvent selection
radio interference
recommended pH range
hydraulic path without solvent selection valve
main assemblies, overview
removing the pump head
repair procedures
replacing firmware
replacing firmware
ignition without cover
physical specifications
restart without cover
results, pressure test
minimum flow rate
power consumption
missing pressure reading
power supply indicator
power consideration
communication settings
running the leak test
motor-drive power
running the pressure test
pressure above upper limit
pressure below lower limit
non-operating altitude
pressure sensor readings
pressure signal missing
safety information
lithium batteries
pressure pulsation
operating Altitude
general information
operating temperature
pressure, operating range
stack configuration
seal wear counter
pump head missing
alternative material
alternative material
for normal phase solvents
parts identification
selection-valve fuse
servo restart failed
1260 Binary Pump VL User Manual
setable flow range
temperature limit exceeded
temperature out of range
simple repair procedures
temperature sensor
site requirements
status indicators
unpacking the pump
solvent selection valve
solvent zero counter
vacuum degasser, when to use
solvent inlet filter
vacuum degasser, operational hints
solvent selection valve
special interfaces
solvent selection
variable reluctance motor
variable stroke volume
forced cold start
stack configuration
electrical and electronic
system setup and installation
optimizing stack configuration
1260 Binary Pump VL User Manual
In This Book
This manual contains technical reference information about the Agilent 1260 Infinity Binary Pump VL (G1312C). The manual describes the following:
• configuring the pump,
• using the pump,
• optimizing performance,
• troubleshooting and diagnostics,
• parts and materials,
• identifying cables,
• hardware information
• safety.
Agilent Technologies 2008, 2010-2012
Printed in Germany 02/2012
Source: https://www.ozhat.com.tr/GetFile.aspx?ID=506588
3rd World Conference on Applied Sciences, Engineering & Technology 27-29 September 2014, Kathmandu, Nepal Environmental Risks of Some Non-steroidal Anti-inflammatory Drugs (NSAIDs) in Surface Water in Ho Chi Minh City O VU HOANG ANH , BUI QUANG MINH , PHAM HONG NHAT 1Institute of Environmental Technology, Vietnam Academy of Science and Technology, Vietnam
CAFFEINE FACT SHEET: CAFFEINE AND HEALTH NTERNATIONAL FOOD INFORMATION COUNCIL FOUNDATION Many women, especially those of childbearing age, are concerned about con-suming too much caffeine. From reproductive effects to osteoporosis, scientists worldwide have studied the effects of caffeine on women's health. This fact sheet reviews the latest research on caffeine and women's health, summarizing the major findings. When taken together, the collective research supports moderate consumption of caffeine (up to 300 mg/day or three cups of coffee) for women, including those who are pregnant or breastfeeding.