Bild 

In Vitro - Ex Vivo Correlation of 
Fluticasone Propionate Pharmacokinetic Profiles 
Maria Börjel1,2, Ewa Selg2 and Per Gerde1,2 ([email protected]) 
1 Inhalation Sciences Sweden AB, Stockholm, Sweden; 2 Institute of Environmental Medicine, Karolinska Institutet, Stockholm, Sweden 
INTRODUCTION 
 
In inhalation drug development there is a need not only for traditional particle 
Analytical quantitation of FP in all samples from the DissolvIt® and IPL experiments was 
The key exposure and pharmacokinetic parameters for the IPL and DissolvIt® system 
characterization data such as size distribution and physicochemical data on the 
performed by LC/MS/MS, with a LLOQ of 100 pg/ml. 
are presented in Table 1. 
formulations, but also for in vitro methods producing data more indicative of subsequent 
in vivo pharmacokinetic behavior. Here we have compared the pharmacokinetic profiles of the low-solubility inhalation steroid fluticasone propionate (FP) in the promising in 
FP (Flixotide) 
FP (Flutiform) 
vitro dissolution/absorption method DissolvIt® (DS) [1] with the well-established 
Parameter 
experimental model for lung pharmacokinetic studies, the ex vivo isolated perfused and 
The concentration of absorbed FP in the perfusate/blood simulant over time is shown in 
ventilated rat lung (IPL) [2, 3]. 
Figure 2A (Flixotide) and 2B (Flutiform). In Figure 2C (Flixotide) and 2D (Flutiform) for both 
models the amount of FP cleared with the perfusate during each sample interval has been 
Exp. duration, (min) 
expressed as percent of the initially deposited dose clearing per second. This 
compensates both for the differing deposited doses of the two exposure models and for 
Tot. Deposited dose, (µg) 
the difference between the varying perfusate flow rate of the IPL and the constant flow rate of the DissolvIt® system. In Figure 3A (Flixotide) and 3B (Flutiform) is shown the amount of 
FP Deposited dose, (µg) 
FP still retained in the rat lung and the DissolvIt® lung simulant model at each sample time, expressed as fraction of the initially deposited dose. The retained FP fraction represents 
, perfusate, (ng/mL) 
either undissolved particles or dissolved substance not yet absorbed by the perfusate buffer/blood simulant. 
Figure 1. The particle size distributions of the generated aerosols for the pMDI formulations containing FP: 
Flixotide (1A) and Flutiform (1B). 
Fraction retained 
(120 min)
Table 1. Key exposure and pharmacokinetic parameters (± standard deviation) for FP (Flixotide and 
Two marketed formulations of FP were tested: Flixotide (GSK), (FP, strength 50 µg) and 
Flutiform) in the IPL and DissolvIt® system. C
: Maximum concentration of drug in the perfusate; 
Flutiform (Mundipharma), (FP, strength 250 µg and formoterol strength 10 µg) provided 
FP (Flixotide) IPL, mean+SD (n=3)
FP (Flutiform) IPL, mean+SD (n=3)
FP (Flutiform) DS, mean+SD (n=3)
: Time to maximum concentration in the perfusate; Fraction perf
: Peak values for the fraction of the 
FP (Flixotide) DS, mean+SD (n=3)
as pressurized metered dose inhalers (pMDIs). The canisters were connected to the US 
deposited dose clearing with the perfusate per second; t
: Time at which Fraction perf
Pharmacopeia Induction Port No 1 of the PreciseInhale® aerosol system and actuated 
Fraction retained 
: Fraction of deposited FP dose left in the DissolvIt® lung simulant and in the rat 
(120 min)
into an air flow of 15 L/min. The aerosols were collected in the PreciseInhale® aerosol 
lung after the 120 min perfusion period; t
: Estimated half-time of drug clearance with the perfusate after 
fitting the fraction retained values to a first order decay curve. 
holding chamber and immediately dispensed to the DissolvIt® and IPL modules at flow 
rates of 1000 mL/min and 250 mL/min, respectively. 
DISCUSSION AND CONCLUSIONS 
Particle size distributions of the generated aerosols (Figure 1) were measured with 
an 8-stage Marple cascade impactor at a flow rate of 2 L/min. 
Previously published methods on in vitro dissolution testing in inhalation drug 
development, such as the flow through cell, paddle apparatus and the Franz cell, all give 
At the start of the DissolvIt® experiment, the aerosol particles deposited on the cover slip 
cumulative dissolution/absorption curves [5, 6] which are not easily comparable to in vivo 
(with PreciseInhale®) are brought into contact with the mucus consisting of 1.5% 
pharmacokinetic profiles. In contrast, the DissolvIt® in vitro dissolution/absorption method 
polyethylene oxide [4] and 0.4% L-alphaphosphatidyl choline (Sigma). The mucus 
FP (Flixotide) IPL, mean+SD (n=3)
FP (Flutiform) IPL, mean+SD (n=3)
FP (Flixotide) DS, mean+SD (n=3)
FP (Flutiform) DS, mean+SD (n=3)
[1] has a dynamic flow-past perfusion strategy which generates concentration curves 
simulant had been applied to a polycarbonate membrane corresponding to the basal 
 values. These values can be more readily compared with human 
membrane of the airway mucosa. The mucus simulant together with the polycarbonate 
Figure 2. Two different scales are used on the y axes: black for IPL values and pink for DissolvIt® values. The 
clinical pharmacokinetic profiles. Here, the rat IPL, previously shown to be a good 
membrane constitutes the diffusion barrier. On the other side of the membrane, the blood 
concentrations of FP from the two formulations: Flixotide (A) and Flutiform (B) in the perfusate/blood simulant 
pharmacokinetic model [2, 3], was used for comparison. 
over time in the IPL (black curves) and the DissolvIt® (pink curves). Panels C (Flixotide) and D (Flutiform) show 
simulant (perfusate consisting of phosphate buffer with 4% albumin) is streaming. 
the percent of deposited doses (IPL black curves and DissolvIt® pink curves) clearing with the perfusate per 
Dissolved particles were absorbed at a perfusate flow rate of 0.4 ml/min. Dissolution was 
In Figure 2 it is shown that the profiles in the DissolvIt® are very similar to the IPL profiles, 
studied by observing particle disappearance using optical microscopy, and by chemical 
especially for Flixotide. For the normalized perfusate clearance values, the curves are also 
analysis of substance removed by absorption in the flow-past perfusate. The amount of 
quite similar in shape. The curves on fraction retained (Figure 3) in the IPL and DissolvIt® 
FP retained in the system (comparable with the amount of drug retained in lung tissue in 
models rank both FP formulations similarly. A large fraction of drug is retained in the 
IPL) was analysed at the end of the 2 h experiments. 
air/perfusate barriers of both systems, as expected from the low solubility of the FP 
formulations. One factor contributing to the faster clearance of Flixotide compared to 
Flutiform in both exposure models, is likely to be the finer particle size distribution of 
The IPL was prepared as previously described [2]. Briefly, whole lungs were isolated from 
Flixotide. Clearance with the perfusate from the IPL is faster than from the DS system for 
female CD IGS (Sprague Dawley) rats (Charles River, Sulzfeld, Germany). The lungs 
both formulations, which is evident both from the steeper declining retention curves 
were ventilated with a negative alternating pressure and perfused in single-pass mode 
(Figure 3) and the nearly doubled normalized perfusate clearance rate of the IPL, as 
with a Krebs-Henseleit buffer containing 4% albumin. For the exposures of the IPL, the 
shown in Figures 2C and D. The main reason for this difference is likely to be the thicker 
FP aerosols actuated via the induction port, were delivered to the lungs (n=3 per 
60 µm diffusion barrier of the DS compared to the air/blood barrier of the IPL being 
formulation) by the PreciseInhale® active dosing system, which calculates in real time the 
dominated by the µm-thick barrier in the alveolar region. 
cumulative inhaled aerosol dose. The system automatically terminates the exposure 
when the inhaled target dose is reached. The liquid propellant evaporated within 
FP single (Flixotide) IPL, mean+SD (n=3)
FP (FP+FFD/ Flutiform) IPL, mean+SD (n=3)
FP single (Flixotide) DS, mean+SD (n=3)
FP (FP+FFD/ Flutiform) DS, mean+SD (n=3)
It is evident, however, that DissolvIt® can generate pharmacokinetic profiles of FP that 
PreciseInhale® system, and thus did not reach the lung. The perfusate was repeatedly 
resemble those in the rat lung. This indicates that DissolvIt® may be a valuable in vitro 
sampled in an automatic fraction collector during a 2 h period post exposure. Thereafter, 
Figure 3. The amount of FP from the two formulations: Flixotide (A) and Flutiform (B), as normalized to the 
dissolution/absorption method to use for IV-IVC in the development of new and generic 
the lungs and trachea were harvested for analysis of the amount of FP retained in the 
initially deposited dose, retained in the rat lung (black curves) and in the DissolvIt® model (pink curves) over 
tissues after the perfusion period. 
 
1. Börjel M, S.R., Gerde P. The DissolvIt: An In Vitro Evaluation of the Dissolution and Absorption of Three Inhaled Dry Powder Drugs in the Lung. in Respiratory Drug Delivery. 2014. Puerto Rico. 2. Selg, E., F. Acevedo, R. Nybom, B. Blomgren, Å. Ryrfeldt, and P. Gerde, Delivering Horseradish Peroxidase as a Respirable Powder to the Isolated, Perfused and Ventilated Lung of the Rat: The Pulmonary Disposition of an Inhaled Model Biopharmaceutical. Journal of Aerosol Medicine and Pulmonary Drug Delivery, 2010. 23(5): p. 273-284. 3. Selg, E., P. Ewing, F. Acevedo, C.O. 
Sjoberg, A. Ryrfeldt, and P. Gerde, Dry powder inhalation exposures of the endotracheally intubated rat lung, ex vivo and in vivo: the pulmonary pharmacokinetics of fluticasone furoate. J Aerosol Med Pulm Drug Deliv, 2013. 26(4): p. 181-9. 4. Shah, S., K. Fung, S. Brim, and B.K. Rubin, An in vitro evaluation of the effectiveness of endotracheal suction catheters. Chest, 2005. 128(5): p. 3699-704. 5. May, S., B. Jensen, M. Wolkenhauer, M. Schneider, and C.M. Lehr, Dissolution techniques for in vitro testing of dry powders for inhalation. Pharm Res, 2012. 29(8): p. 2157-66. 6. 
Salama, R.O., D. Traini, H.K. Chan, and P.M. Young, Preparation and characterisation of controlled release co-spray dried drug-polymer microparticles for inhalation 2: evaluation of in vitro release profiling methodologies for controlled release respiratory aerosols. Eur J Pharm Biopharm, 2008. 70(1): p.145-52. 
Source: http://www.inhalation.se/wp-content/uploads/2015/12/DDL2015-Poster.pdf
   Pharmacology of Systemic Antibacterial Agents:  Leena Palomo, DDS, MSD; Géza T. Terézhalmy, DDS, MA  Continuing Education Units: 3 hours Online Course:  Disclaimer: Participants must always be aware of the hazards of using limited knowledge in integrating new techniques or  procedures into their practice. Only sound evidence-based dentistry should be used in patient therapy.
    Cell, Vol. 103, 1121–1131, December 22, 2000, Copyright 2000 by Cell Press Human Upf Proteins Target an mRNA for Nonsense-Mediated Decay When BoundDownstream of a Termination Codon Jens Lykke-Andersen, Mei-Di Shu, immunoglobulin and T cell receptor genes undergo so- and Joan A. Steitz*